Emerson Rosemount 3051SAL Reference Manual page 130

Table of Contents

Advertisement

Appendix A: Specifications and reference data
May 2013
Table A-9. Rosemount 3051SFC Compact Orifice Flowmeter Ordering Information
The Standard offering represents the most common models and options. These options should be selected for best delivery.
__The Expanded offering is subject to additional delivery lead time.
Special Configuration (Software)
Standard
(20)
C4
NAMUR alarm and saturation signal levels, high alarm
(20)(20)
C5
NAMUR alarm and saturation signal levels, low alarm
(20)(20)
C6
Custom alarm and saturation signal levels, high alarm
Note: Requires option code C1, custom software configuration. A Configuration Data Sheet must be
completed, see document number 00806-0100-4810.
(20)(20)
C7
Custom alarm and saturation signal levels, low alarm
Note: Requires option code C1, custom software configuration. A Configuration Data Sheet must be
completed, see document number 00806-0100-4810.
(20)(20)
C8
Low alarm (standard Rosemount alarm and saturation signal levels)
Special Configuration (Hardware)
Standard
(20)(26)
D1
Hardware Adjustment (zero, span, security).
D4
External ground screw
(20)(26)
DA
Hardware adjustment (zero, span, security) and external ground screw
Conduit Electrical Connector
Expanded
(31)
ZE
M12, 4-pin, Male Connector (eurofast)
(31)
ZM
A size Mini, 4-pin, Male Connector (minifast)
(1) Not available for Primary Element Type code C.
(2) For 2-in. (50.8 mm) line sizes the Beta Ratio is 0.6 for Primary Element Type code C.
(3) Option code R is required for Measurement Type codes 1, 3, 5, and 7.
(4) Required for Measurement Type codes 2, 4, 6, and D.
(5) Required for Measurement Type codes 3, 4, 7, and D.
(6) For Measurement Type 1, 2, 5, and 6 with DP range 1, absolute limits are 0.5 to 2000 psi (0,03 to 137,9 bar) and gage limits are
-14.2 to 2000 psig (-0,98 to 137,9 bar).
(7) Requires PlantWeb housing.
(8) Only available with Measurement Type D.
(9) Available approvals are FM Intrinsically Safe, Division 2 (option code I5), CSA Intrinsically Safe (option code I6), ATEX Intrinsic Safety
(option code I1; only available with 2.4 GHz), and IECEx Intrinsic Safety (option code I7; only available with 2.4 GHz).
(10) Material specified is cast as follows: CF-8M is the cast version of 316 SST, CF-3M is the cast version of 316L SST. For housing, material is aluminum with polyurethane paint.
(11) Available with output code A only. Available approvals are FM Intrinsically Safe, Division 2 (option code I5), ATEX Intrinsic Safety (option code I1), or IECEx Intrinsic Safety
(option code I7). Contact an Emerson Process Management representative for additional information.
(12) Available with output code A only. Not available with approvals. See Rosemount 753R Product Data Sheet, 00813-0100-4379, to specify Web-Based Monitoring Indicator.
Does not integrate into plant host systems.
(13) Not available with Wireless Operating Frequency and Protocol option codes 1 or 2.
(14) Not available with Wireless Operating Frequency and Protocol option codes 1 or 2 or Housing code 01. This option is only available with differential pressure ranges 2 and 3,
and silicone fill fluid.
(15) Not available with Primary Element Type code P.
(16) Not available with Output Protocol code F or Housing code 01.
(17) Only available with Measurement Type D.
(18) Not available with output code F. Not available with housing code 01 or 7J.
(19) Instrument valves not included in Traceability Certification.
(20) Materials of Construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental limits apply to certain
materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
(21) Only available with Measurement Types 1-7.
(22) Not available with M20 or G ½ conduit entry size.
(23) Not available with Housing code 01 or 7J.
(24) Not available with output code F, Housing code 01, option code DA1, or option code QT.
(25) Cable supplied is Belden 3084A, rated for ambient temperatures up to 167°F (75°C).
(26) Not available with Housing code 00, 01, 5A, or 7J.
(27) Available with Output Protocol code A and Plantweb housing only.
(28) Requires PlantWeb housing and output code F.
(29) Requires PlantWeb housing and output code A. Includes Hardware Adjustments as standard.
(30) Requires Rosemount Engineering Assistant version 5.5.1 to configure.
(31) Not available with Housing code 00, 01, 5A, or 7J. Available with Intrinsically Safe approvals only. For FM Intrinsically Safe, Division 2 (option code I5) or
FM FISCO Intrinsically Safe (option code IE), install in accordance with Rosemount drawing 03151-1009 to maintain outdoor rating (NEMA 4X and IP66).
122
Reference Manual
00809-0200-4802, Rev BA
Specifications and Reference Data
Standard
Standard

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3051shpRosemount 3051s

Table of Contents