Rigol 2000 Series User Manual page 178

Digital oscilloscope
Hide thumbs Also See for 2000 Series:
Table of Contents

Advertisement

RIGOL
or cursor B (only channels enabled are available). You can also select "None",
namely do not use cursor A or cursor B.
2. Adjust Cursor Position
Adjust cursor A: press CursorA and use
A. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor B: press CursorB and use
B. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor A and B at the same time: press CursorAB and use
adjust the position of cursor A and B at the same time. During the
adjustment, the measurement results will change accordingly. The
adjustable range is limited within the screen.
Note:
Under the current menu, you can press
CursorB and CursorAB.
When "None" is selected as the measurement source of cursor A (or cursor
B), CursorA (or CursorB) and CursorAB will be grayed out.
In track mode, the cursor will track the marked point (namely jumps up and
down with the transient change of the waveform). Thus, the Y value might
change even though you do not adjust the cursor.
3. Measurement Example
Use cursor A and B to measure the waveforms of CH1 and CH2 respectively. You
would discover that the cursors would track the marked points when the
waveforms are expanded or compressed horizontally, as shown in the figures on
the next page.
6-36
Chapter 6 MATH and Measurements
to adjust the position of cursor
to adjust the position of cursor
continuously to select CursorA,
MSO2000A/DS2000A User Guide
to