Chapter 1 Quick Start
Note:
The digital channel input terminal does not support hot plugging. Please do
not insert or pull out the logic probe when the instrument is in power-on state.
Function Inspection
Press Storage Default to restore the oscilloscope to its factory states.
1.
2.
Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in the figure below.
3.
Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Probe Compensation Signal Output Terminal".
Figure 1-7 Probe Compensation Signal Output Terminal/Ground Terminal
Set the attenuation on the probe to 10X. Then press AUTO.
4.
5.
Observe the waveform on the display. If the waveforms actually shown do not
match that in the figure below, please perform "Probe Compensation".
6.
Use the same method to test the other channels.
WARNING
To avoid electric shock during the use of probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
MSO2000A/DS2000A User Guide
Figure 1-8 Probe Compensation Signal
Probe Compensation Signal Output Terminal
Ground Terminal
RIGOL
1-7