Leak Sensor Short

Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak
is detected when solvent cools the leak sensor, causing the leak sensor
current to change within defined limits. If the current increases above the
upper limit, the error message is generated.
Probable cause
1
2

Compensation Sensor Open

Error ID: 0081
The ambient- compensation sensor (NTC) on the main board in the module
has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient
temperature changes. If the resistance across the sensor increases above
the upper limit, the error message is generated.
Probable cause
1
Agilent 1260 Infinity VWD User Manual
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Defective main board.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.