Problem / Error
Barcode reading
failure
Data Integrity Problem An error has occurred
Instrument internal
communication
problem
Instrument door
problem
M01534_v1.1 MagicPrep NGS Instrument Operating Manual
Possible Cause
An error has occurred
during the barcode
reading process.
while downloading log
files.
An error has occurred
within the library
preparation process.
An error has occurred
within the library
preparation process,
possibly due to the front
door being opened.
Corrective Measure
If the error still exists, please contact Tecan
Customer Support.
Follow the on-screen instructions to remove
the reagent cartridge and sample deck.
 Switch off the instrument
 Switch on the instrument to initiate the self-
test/diagnostic sequence.
 Reload the MagicPrep NGS and try to
repeat the library preparation.
If the error still exists, please contact Tecan
Customer Support.
Remove and reinsert the USB drive and retry
downloading the log files.
If the problem is not resolved, check if a new
SW update is available
If you are operating on the latest released SW
version
 Switch off the instrument
 Switch on the instrument to initiate the self-
test/diagnostic sequence.
If the error still exists, please contact Tecan
Customer Support.
Stop the library run
Follow the on-screen instructions to remove
the used reagent cartridge and sample deck
Check if a new SW update is available
If you are operating on the latest released SW
version
 Switch off the instrument
 Switch on the instrument to initiate the self-
test/diagnostic sequence.
 Reload the MagicPrep NGS and try to
repeat the library preparation.
If the error still exists, please contact Tecan
Customer Support.
Stop the library run
Follow the on-screen instructions to remove
the used reagent cartridge and sample deck
Check if a new SW update is available
Operating Manual
Page 31 of 39