Emission And Immunity - tecan MagicPrep NGS Operating Manual

Table of Contents

Advertisement

Operating
conditions
The MagicPrep NGS instrument has been designed to be able to operate under the above
environmental conditions. However, with MagicPrep NGS reagent kits, due to the presence of
temperature sensitive enzymes and specific temperature dependent reactions during the NGS
library preparation workflow, narrower operating conditions are required.
Operating
conditions for
NGS library
preparation
Instrument
transport
conditions
Instrument
storage
conditions
3.6

Emission and Immunity

Noise emission <60 dBA (sound pressure), measured at a distance of 1 meter from the
instrument
EMC The MagicPrep NGS instrument has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the FCC Rules and
CISPR 11/EN 55011. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which
case the user will be required to correct the interference at his own expense.
It is the operator's responsibility to ensure that a compatible electromagnetic
environment for the MagicPrep NGS can be maintained in order that the
MagicPrep NGS will perform as intended. Do not operate the MagicPrep NGS in
close proximity to sources of strong electromagnetic radiation (e.g., unshielded
intentional RF sources), as these can interfere with the proper operation.
M01534_v1.1 MagicPrep NGS Instrument Operating Manual
Operating Temperature
Operating humidity
Operating altitude
Room temperature
Operating humidity
Operating altitude
Transport temperature
Transport humidity
Storage temperature
Storage humidity
15-32°C (59-90°F)
30-80% relative (non-condensing) at 30°C
(86°F)
Max. 2000 m above sea level
20-26°C (68-79°F)
30-60% relative (non-condensing)
About 500 m above sea level
-20 to 60°C (-4 to 140 °F)
20-80 % relative (non-condensing)
1-60°C (34-140 °F)
5-80% relative (non-condensing) at 30°C (86
°F) or below
Operating Manual
Page 15 of 39