Calibration options
Check and adjust
step-by-step
86
3.
Select:
Hz Collimation, or
•
V-Index.
•
Compensator
•
View Calibration.
•
In the Check & Adjust screen there are several calibration options.
Menu selection
Hz Collimation
V-Index
Compensator
View Calibration
1.
Level the instrument with the electronic level. Refer to
Level up with the electronic level
2.
0022423_001
3.
Press OK to measure to the target point.
4.
180°
0022424_001
☞
For checking the horizontal aim, the difference in Hz and
V are displayed.
5.
Press OK to measure to the target point.
☞
The old and new calculated values are displayed.
6.
Either:
Press OK to save the new adjustment data, or
•
Press
to exit without saving the new adjustment data.
•
Description
Refer to
Check and adjust
Refer to
Check and adjust
Refer to
11.1.4 Adjusting the
Displays the current calibration values and
compensator indexes that have been set for
HA-Collimation and V-index.
step-by-step.
Aim at a point approximately 100
m from the instrument which is
within 5° of the horizontal.
± 5°
Change face and aim at the tar-
get point again
180°
step-by-step.
step-by-step.
Compensator.
4 Operation
-
Tools