4.0
Calibration
The
can be calibrated using the
Tracer AV
utility. Each method consists of the following steps:
•
Zero calibration
•
Entering the test weight value
DEFCAL
You can also use the default calibration feature, which utilizes a more user-friendly procedure; however, not all
options are available in default calibration.
4.1
Default Calibration
If you use a 500lb or 300lb scale, you can select it
from the
menu. This automatically uses the
DEFCAL
settings shown in Table 4-1. for 500lbs or Table 4-2.
for 300lbs.
Primary
Secondary
SP1 Trip
SP2
Digital Filter
Zero Tracking
Table 4-1. 500lb scale settings
Primary
Secondary
SP1 Trip
SP2
Digital Filter
Zero Tracking
Table 4-2. 300lb scale settings
and
Zero
Units
CONFIG
XXXXXXX
FORMAT
CALIBR
WZERO
WV AL
Figure 4-1. Calibration (CALIBR) Menu
XXXXXXX
DEFCAL
CONFIG
XXXXXXX
FORMAT
SELECT
WZERO
500lb
300lb
Figure 4-2. Default Calibration (DEFCAL) Menu
500lb x 1lb
250kg x .5kg
50lb
Off
8,8,8
3D
300lb x 1lb
150kg x .5kg
50lb
Off
8,8,8
3D
keys, EDP commands, or the
•
Span calibration
•
Optional rezero calibration for test weights
using hooks or other additional items.
SERIAL
PROGRM
XXXXXXX
SETPNT
WSPAN
REZERO
Zero
CALIBR
SERIAL
PROGRM
XXXXXXX
SETPNT
WV AL
WSPAN
To calibrate the
Tracer AV
keys, do the following:
1. Place the
Tracer AV
To enter setup mode, remove the setup switch
access screw on the
4). Once the screw is removed, use a small
screwdriver to press the setup switch. The
display changes to show the word
2. Remove all weight from the scale platform. If
your test weights require hooks or other
items, place them on the scale.
3. Press the
Zero
4. Press the
Units
(
).
WZERO
5. Press the
Zero
indicator displays
progress. When complete, the A/D count for
the zero calibration is displayed. DO NOT
adjust this value after WSPAN has been set!
Press
Zero
value and go to the next prompt (
Tracer AV Installation Manual - Calibration
®
configuration
Revolution
VERS
VERS
using the
and
Zero
in setup mode
(See Figure 2-3
Tracer AV
DEFCAL
key to go to
.
SELECT
key to go to zero calibration
key to calibrate zero. The
while calibration is in
*CAL*
again to save the zero calibration
WVAL
Units
.
).
23