Healthweigh® Indicators

Adjusting the Power Down Time Limit

Product: Healthweigh Indicator
Issue: By default, the indicator automatically powers down after 9 minutes when powered with batteries only
Solution: Adjust the TOFF setting
When the healthweigh indicator is powered by batteries only, it powers down automatically after 9 minutes. Power down time
can be configured from 1-9 minutes or disabled.
Manuals and additional resources are available from Rice Lake Weighing Systems at
Warranty information is available at
Use the following procedure to adjust the display time out setting:
1. Remove the four screws securing the back cover to the indicator (see
2. Press and hold the ON/OFF and ENTER keys until the ID displays.
June 6, 2023
www.ricelake.com/warranties
Screws
Back Cover
Figure 1. Back Cover Screw Locations
www.ricelake.com/manuals
Figure
1).
PN 219265 Rev A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Healthweigh and is the answer not in the manual?

Questions and answers

Summary of Contents for Rice Lake Healthweigh

  • Page 1 Issue: By default, the indicator automatically powers down after 9 minutes when powered with batteries only Solution: Adjust the TOFF setting When the healthweigh indicator is powered by batteries only, it powers down automatically after 9 minutes. Power down time can be configured from 1-9 minutes or disabled.
  • Page 2 IMPORTANT: If value is set to 0, then it stays powered on until the ON/OFF key is pressed to power down the scale. © Rice Lake Weighing Systems Content subject to change without notice. 230 W. Coleman St. • Rice Lake, WI 54868 • USA USA: 800-472-6703 • International: +1-715-234-9171 PN 219265 Rev A June 6, 2023 www.ricelake.com...

Table of Contents