3.2.2
Configuration Menu
DEFCAL
CONFIG
GRADS
1000
number
DIGFL1
8
16
32
64
1
2
4
CONFIG Menu
Parameter
Choices
Level 2 submenus
GRADS
1000
number
ZTRKBN
0
number
ZRANGE
1.900000
number
FORMAT
CALIBR
SERIAL
XXXXXXX
ZTRKBN
ZRANGE
0
1.900000
number
number
DIGFL2
DIGFL3
8
16
32
64
1
2
4
Figure 3-5. Configuration Menu
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.
Display divisions for primary and secondary units are specified on the FORMAT menu.
Automatically zeroes the scale when within the range specified, as long as the input is
within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.
NOTE: For scales using linear calibration, do not set the zero tracking band to a value
greater than that specified for the first linearization point.
Selects the range within which the scale can be zeroed. The 1.900000 default value is ±
1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use the default value for legal-for-trade applications.
Table 3-4. Configuration Menu Parameters
PROGRM
SETPNT
VERS
XXXXXXX
MOTBAN
1
number
DFSENS
8
8OUT
16
16OUT
32
32OUT
64
64OUT
1
128OUT
2
2OUT
4
4OUT
Description
Tracer AV Installation Manual - Configuration
OVRLOA
SMPRAT
FS+2%
15HZ
FS+1D
30HZ
FS+9D
60HZ
FS
7.5HZ
DFTHRH
T AREFN
NONE
BOTH
2DD
NOT ARE
5DD
PBTARE
10DD
KEYED
20DD
50DD
100DD
200DD
250DD
13