Siemens MAMMOMAT Novation DR Installation Manual page 78

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

9 - 18
7.
Fig. 19 Acquisition System Menu
NOTE
8.
9.
10.
11.
12.
13.
DR
MAMMOMAT Novation
System
If you want to end the Bad Pixel Mapping operation at this point,
click the "Cancel" button.
Set the following exposure parameter on the control console of the MAMMOMAT
DR
Novation
:
Mo/Mo, 28KV, 250mAs
Insert the collimator mounted plexi phantom.
Mount the compression plate simulator.
Double click on the first Image of the 8 at the Procedure: Calibration Procedure
Start an exposure.
Wait until the image is loaded into the viewer. The following lines are displayed in
the message bar.
Image acquisition is in progress .....
Saving processed Image .....
Loading Image ......
Click the Accept or Reject button.
SPB7-250.815.02
Rev. 02
09.04
Image Quality
Page 18 of 20
CS SD 24
Siemens AG
Medical Solutions