Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Grid object table • Mammography bar pattern (5-20 lp/mm), for instance Nuclear Associates Resolution Test Pattern (07-555), http://www.nucl.com • Cleaning agents* specified in Instructions for Use MAMMOMAT 1000/3000 Nova , chapter Cleaning, Disinfection • Standard service tools* Siemens AG Register 8 SPB7-230.101.03...
Tests or adjustments that must be made with radiation are identified with the radiation warning symbol During these types of adjustments, radiation protective clothing must be worn. Opdimar Register 8 SPB7-230.101.03 Page 2 of 2 Siemens AG Maintenance Instructions Rev. 04 07.03 CS PS 24 Medical Solutions...
Opdima r , chapter Instruction for obtaining the systems host ID and hostname . • Document country settings in the Maintenance Protocol. • If applicable, document network and printer parameters in the Maintenance Protocol. Siemens AG Register 8 SPB7-230.101.03 Page 1 of 2 Opdimar Medical Solutions Rev.
Page 8
2 - 2 System Status This page intentionally left blank. Opdimar Register 8 SPB7-230.101.03 Page 2 of 2 Siemens AG Maintenance Instructions Rev. 04 07.03 CS PS 24 Medical Solutions...
Page 9
Check that the unit is installed in a clean environment and is not exposed to dust. • Check that right MO media is used. Only Sony standard media (Siemens part No. 66 33 734, Sony RMO-S561, 9.1 GB rewritable) shall be used as MO media. •...
Page 10
• Check that the biopsy unit is safely locked to the MAMMOMAT. Opdimar Register 8 SPB7-230.101.03 Page 2 of 2 Siemens AG Maintenance Instructions Rev. 04 07.03 CS PS 24 Medical Solutions...
For instructions how to check Opdimar AEC see Service Instructions for Opdima chapter Check of Opdima AEC . Resolution For instructions how to check resolution see Service Instructions for Opdima , chapter Check of resolution . Siemens AG Register 8 SPB7-230.101.03 Page 1 of 2 Opdimar Medical Solutions Rev.
Page 12
4 - 2 Testing This page intentionally left blank. Opdimar Register 8 SPB7-230.101.03 Page 2 of 2 Siemens AG Maintenance Instructions Rev. 04 07.03 CS PS 24 Medical Solutions...
Clean the unit using the materials recommended in Instructions for use for MAMMOMAT 1000/3000 Nova (chapter Cleaning, Disinfection ) and repair any damaged paint. • System ON. Final test exposure Take one final test exposure. Siemens AG Register 8 SPB7-230.101.03 Page 1 of 2 Opdimar Medical Solutions Rev.
Page 14
5 - 2 Miscellaneous This page intentionally left blank. Opdimar Register 8 SPB7-230.101.03 Page 2 of 2 Siemens AG Maintenance Instructions Rev. 04 07.03 CS PS 24 Medical Solutions...