Siemens MAMMOMAT Novation DR Installation Manual page 23

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Detector installation
3.
4.
5.
CAUTION
6.
Fig. 5 Check temperature logger.
NOTE
NOTE
Siemens AG
Medical Solutions
Open box, remove packing insert, temperature logger and phase change material.
Remove the packing material inside the inner box, which includes the temperature
logger that is used to ensure that the detector maintained the proper temperature
while being shipped and stored.
Carefully remove the detector from its protective bag.
Handle the detector with extreme care!
Avoid touching the thin film top cover that protects the imaging
area. The imaging electronics under the front cover are extremely
sensitive. The actual image area starts at the front edge and is
approximately 25 mm inside the other three edges.
Any dust or fingerprints in this area may affect final image quality.
Temperature extremes may cause permanent damage.
Check the temperature logger with a ball pen, as shown in (Fig. 5).
When the LED lights green, transportation was OK and the detec-
tor can be installed.
When the LED lights red, the detector is defective and has to be
replaced.
You have to send the temperature logger to the factory using the
regular Customer Services return process.
You'll find all documents required for sending it enclosed in the
detector package (black box).
Enter the detector serial number, your district address and instal-
lation date before you sen it.
SPB7-250.815.02
Rev. 02
09.04
MAMMOMAT Novation
Page 3 of 6
CS SD 24
4 - 3
DR
System

Advertisement

Table of Contents
loading

Table of Contents