Image Quality
NOTE
14.
15.
16.
Siemens AG
Medical Solutions
If you Reject the image, a new image can be acquired for the Gain Calibration.
Correct the problem select the same image from the Procedure: Calibration
Procedure and continue with step 12.
You would "Reject" an image if it exhibits any edge cut-off due to
collimation or misalignment or if there are any artifacts from debris
or obstructions.
Wait approximately 10 seconds for the image to be calculated (observe the status
bar) and then double click on the next image in the Procedure: Calibration
Procedure and proceed with step 12 until you have a total of 8 images.
Click Done to exit the Calibration.
Close the examination.
SPB7-250.815.02
Rev. 02
09.04
MAMMOMAT Novation
Page 13 of 20
CS SD 24
9 - 13
DR
System