Collimator Wing Difference Calibration - Siemens MAMMOMAT Novation DR Installation Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Collimator Adjustment
10.
11.
12.

Collimator Wing Difference Calibration

The collimator wing difference calibration has to be set to 0.
Fig. 4 Collimator wing difference calibration
1.
2.
Siemens AG
Medical Solutions
Continue with Step 7 until the beam field and light field are identical.
Save the calibration and Exit the Service Software.
Record the measured values in the "Startup Protocol" on Page 11 - 1 .
Start the service software on the Service PC and select:
Configuration -> Collimator -> Calibration of wing differance
Save the calibration and Exit the Service Software.
SPB7-250.815.02
Rev. 02
09.04
MAMMOMAT Novation
Page 5 of 6
CS SD 24
7 - 5
7
DR
System