Siemens MAMMOMAT Novation DR Installation Manual page 89

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Startup Protocol
kV
mAs
Anode/
filter
_____V x 5 kV =
30 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
10 mAs
W/Rh
_____V x 5 kV =
25 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
-
Mo/Mo
_____V x 5 kV =
25 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
-
W/Rh
Tab. 2
Siemens AG
Medical Solutions
Measured values
Tube high voltage, current and mAs values
SPB7-250.815.02
Rev. 02
09.04
Result
kV must be &5%
mAs must be &10%
______ kV
______ mAs
______ kV
______ mAs
Display ______ mAs
______ kV
______ mAs
Display ______ mAs
MAMMOMAT Novation
Page 5 of 10
CS SD 24
11 - 5
DR
System