Siemens MAMMOMAT Novation DR Installation Manual page 36

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

5 - 10
5.
Fig. 11Foot switch
6.
7.
Fig. 12 Display at the lower part of the stand
8.
DR
MAMMOMAT Novation
System
Check the function of the foot switch for motorized compression/decompression
(1 and 2/Fig. 11).
2
Press the foot switch for motorized decompression to move the compression plate
upwards while at the same time pressing the compression plate in the opposite
direction by hand.
The upward movement of the compression plate shall now stop (safety switch).
Set the compression to 7 kg by using the presetting knob (5/Fig. 10).
The preset compression force is shown on the display at the lower part of the
stand (1/Fig. 12).
1
Press the foot switch for motorized compression to move the compression plate
downwards while simultaneously pressing the compression plate in the opposite
direction by hand.
1
SPB7-250.815.02
Rev. 02
09.04
Mechanical Checks
Page 10 of 12
CS SD 24