Siemens MAMMOMAT Novation DR Installation Manual
Siemens MAMMOMAT Novation DR Installation Manual

Siemens MAMMOMAT Novation DR Installation Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Quick Links

MAMMOMAT Novation
System
Startup
with WH AWS
Print No.: SPB7-250.815.01.02.02
Replaces: SPB7-250.815.01.01.02
DR
© Siemens AG 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
English
Doc. Gen. Date: 09.04
SP
reserved.

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MAMMOMAT Novation DR

  • Page 1 MAMMOMAT Novation System Startup with WH AWS © Siemens AG 2004 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Mounting covers ........5 - 1 MAMMOMAT Novation Siemens AG SPB7-250.815.02...
  • Page 4 Checking Grid Lines ........9 - 1 MAMMOMAT Novation SPB7-250.815.02 Page 4 of 6 Siemens AG System Rev. 02 09.04...
  • Page 5 Starting up and using the service PC ..... . . 12 - 6 MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 6 Medical Solutions Rev.
  • Page 6 Troubleshooting PC connection ......12 - 6 13 _____ Changes to previous version ____________________________________13 - 1 MAMMOMAT Novation SPB7-250.815.02 Page 6 of 6 Siemens AG System Rev. 02 09.04 CS SD 24...
  • Page 7: Configuration

    • detector and • syngo Acquisition Workstation. Depending on the installation configuration, some points may be omitted. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 4 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 8: Meters, Appliances And Tools

    A non-invasive digital kV meter If not available in the district, a scope can be used instead Tab. 1 Meters, appliances and tools MAMMOMAT Novation SPB7-250.815.02 Page 2 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 9: Phantoms / Auxiliary

    Standard installation and service tools Torque wrench for bolting the stand/ console to the floor Electrical screwdriver with adjustable torque is recommended Tab. 3 Tools MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 4 Medical Solutions Rev. 02 09.04 CS SD 24 System...
  • Page 10 1 - 4 Prerequisites This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 11: Symbols

    Maintenance measurements must be made according to the DHHS Maintenance Instructions (SPB7-250.662.01...) and the DHHS Supplements to the instructions for use (SPB7-250.661.01...). Results must be recorded in DHHS Measurement certificates (SPB7-250.663.01...). MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 4 Medical Solutions Rev. 02 09.04...
  • Page 12: Protective Measures

    They may cause severe injury. These strips must always be applied onto the edges of the metal curtain during service and maintenance work. MAMMOMAT Novation SPB7-250.815.02 Page 2 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 13: Measurements With The Oscilloscope

    The Acquisition Workstation service software is web-based and can be started from the Options application menu. • BRICK The BRICK service software can be accessed via the Acquisition Workstation. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 4 Medical Solutions Rev. 02 09.04...
  • Page 14: Delay Times Between Two Exposures

    200 min. 30 max. 300 min. 45 max. 400 min. 60 max. 500 min. 75 Tab. 1 Delay times between two exposures MAMMOMAT Novation SPB7-250.815.02 Page 4 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 15: Radiation Protection Door Switch (Optional)

    Connect the cable to the lamp holder in series with the power supply. CAUTION When connecting the exposure lamp to the Master board, make sure that the electrical load does not exceed 24VAC / 25W MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 6 Medical Solutions Rev. 02 09.04...
  • Page 16: Mains Voltage

    Mains Frequency: 50 or 60 Hz Maximum load: 10kVA (momentary) and 0.8 kVA (long time load). Fuse: 25 A (external) 208 -> 400 V. MAMMOMAT Novation SPB7-250.815.02 Page 2 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 17: Connections

    Fig. 3Incoming mains cable Connect the incoming mains cable to the mains terminal (Fig. 3) at the bottom of the back of the stand. Strain-relieve the incoming mains cable. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 6 Medical Solutions Rev. 02 09.04...
  • Page 18: Measuring The Line Resistance

    (generator) are on: LED check 28 VP 5 VF - 15 V + 15 V 28 VS Tab. 2 LED check MAMMOMAT Novation SPB7-250.815.02 Page 4 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 19: Checking The Microprocessors Led's

    V3—Power status. A green LED that lights to indicate that the PXCM is provid- ing power to the Detector. Check that the operating mode dr is displayed on the control console. Check if the noted exposure parameters are restored correctly. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 6 Medical Solutions Rev. 02 09.04...
  • Page 20: Startup Of The Acquisition Workstation

    Rotate the swivel arm system to 0 degrees by pressing one of the switches for clockwise rotation on the supporting arm. System and mains OFF. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 21: Unpacking Of The Detector

    Fig. 2 Opening the box Remove all of the other packing material covering the inner box in which the detector is located. NOTE Do not cut the lifting band. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 6 Medical Solutions Rev. 02 09.04...
  • Page 22 Fig. 4 Remove packing material Slide the lifting band from the inner box. Lift out the detector and carefully remove from wrapper. MAMMOMAT Novation SPB7-250.815.02 Page 2 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 23 You’ll find all documents required for sending it enclosed in the detector package (black box). Enter the detector serial number, your district address and instal- lation date before you sen it. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 6 Medical Solutions Rev. 02 09.04...
  • Page 24: Mounting The Detector

    Connect the detector cable (3/Fig. 6) to the cable coming out of the wing (2/Fig. 6). Mount the cover over the detector cable with 2 screws. Make sure that no cables are squeezed. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 25: Mounting The Grid

    Connect the cables (5/Fig. 7) for the magnification table, according to the labels, on both sides. WARNING Make sure that the cables are connected according to the labels, otherwise the D805 will burn up. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 6 Medical Solutions Rev. 02 09.04...
  • Page 26: Adjusting The Switches

    Mount the top wing cover with 4 screws. Slide the detector fibre cover carefully over the detector and fasten it with 6 screws at the bottom of the cover. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 27: Mounting Covers

    Connect ground wires (3/Fig. 1) to patient handles (4/Fig. 1). Side covers Screws in different sizes are included in the delivery. Fig. 2 Covers and sign MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 12 Medical Solutions Rev. 02 09.04...
  • Page 28: Front Cover

    Risk of damage. Be careful while attaching the front cover. Fasten the front cover to the side covers with two short screws (5/Fig. 2). MAMMOMAT Novation SPB7-250.815.02 Page 2 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 29: Stand Covers

    Fit the rear cover (4/Fig. 3) to the stand and fasten it with screws (5/Fig. 3) and contact washers (6/Fig. 3). NOTE Be careful not to damage the cables at the cable outlet! MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 12 Medical Solutions Rev. 02 09.04...
  • Page 30: Face Shield

    Place the Stand cap on the top of the MAMMOMAT stand (2/Fig. 4). Face shield Fig. 5 Face shield Slide the face shield (1) onto the holder for external diaphragms.Checking the swivel arm system MAMMOMAT Novation SPB7-250.815.02 Page 4 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 31: Rotary Motion

    (+135 degrees clockwise and -180 degrees counterclockwise). The projection angle is shown on the display at the lower part of the stand. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 12 Medical Solutions Rev. 02 09.04...
  • Page 32: Vertical Adjustment

    "Metal Curtain" on Page 2 - 2. Press the buttons for upward and downward movement (3/Fig. 6) of the swivel arm system. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 33: Emergency Stop

    The button will latch and all motorized movements of the equipment will be blocked ("Er 813" shall appear). To reset the emergency stop button, turn it clockwise. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 7 of 12 Medical Solutions Rev. 02 09.04...
  • Page 34: Checking The Field Light Time

    (1/Fig. 9). The field light shall go out automatically after a preset period of time (normally 20 s.). Fig. 9 Foot switch MAMMOMAT Novation SPB7-250.815.02 Page 8 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 35: Adjusting The Field Light Time

    Check the function of the knobs for manual compression/decompression (3/Fig. 10). Place a piece of soft material on the object table, for example a rolled up towel or a piece of foam rubber (4/Fig. 10). MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 9 of 12 Medical Solutions Rev. 02 09.04...
  • Page 36 Press the foot switch for motorized compression to move the compression plate downwards while simultaneously pressing the compression plate in the opposite direction by hand. MAMMOMAT Novation SPB7-250.815.02 Page 10 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 37: Decompression Button

    Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this setting for at least two weeks before further changes are made. OPCOMP has been thoroughly tested by Siemens. If any factors NOTE are changed, OPCOMP will no longer be an optimization of com- pression force and image quality according to Siemens’...
  • Page 38: Checking The Exposure Blocking

    Check that the exposure release is blocked during the following conditions: • No patient is registered. • No compression plate is mounted. • The wing is not in position. MAMMOMAT Novation SPB7-250.815.02 Page 12 of 12 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 39: Tube High Voltage, Current And Mas Values

    Be aware, that with the AEC mode on, a pre-exposure of 50 ms is done. Select the correct trigger on the scope, so you get the pre-expo- sure and the all the exposure times. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 6 Medical Solutions Rev. 02 09.04...
  • Page 40 1 V/T (1 T <=> 40 mA) 132 mA 1 V/T (1 T <=> 40 mA) 0.9 s 0.2 s/T 0.72 s 0.2 s/T MAMMOMAT Novation SPB7-250.815.02 Page 2 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 41: Checks With A Large Focus And Without Aec

    Record the measured kV and mAs value in the "Startup Protocol" on Page 11 - 1 Change the setting and continue with the next measurement until you have completed all measurements from the table above. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 6 Medical Solutions Rev.
  • Page 42: Checks With A Small Focus And Without Aec

    Record the measured kV and mAs value in the Startup Protocol, page 11 - 1. Change the setting and continue with the next measurement until you have completed all measurements from the table above. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 43: Checks With A Large Focus And With Aec

    Record the measured kV and mAs value in the Startup Protocol, page 11 - 1. Change the setting and continue with the next measurement until you have completed all measurements from the table above. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 6 Medical Solutions Rev.
  • Page 44 6 - 6 X-ray tube checks This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 45: Automatic Collimator

    The x-axis is located at the object table chest wall side edge. Fig. 1 Collimator calibrations • Moving the collimator OUT is done by INCREASING the YC value and DECREASING the YS, XL and XR values. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 6 Medical Solutions Rev. 02 09.04...
  • Page 46: Collimator Check

    Collimator calibration is carried out for each focus including the field light focus. Connect the compression plate simulator and the collimator mounted plexi phan- tom. MAMMOMAT Novation SPB7-250.815.02 Page 2 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 47 Measure the white line thickness at each image edge with the application GUI and adjust the corresponding XL, XR, YC and YS values with the service software Start an exposure. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 6 Medical Solutions Rev.
  • Page 48: Collimator Field Light Calibration

    Compare the image with the reading of the X-ray cross on the detector when the light field is turned on. Adjust the difference with the Service Software. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 49: Collimator Wing Difference Calibration

    Fig. 4 Collimator wing difference calibration Start the service software on the Service PC and select: Configuration -> Collimator -> Calibration of wing differance Save the calibration and Exit the Service Software. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 6 Medical Solutions Rev. 02 09.04...
  • Page 50 7 - 6 Collimator Adjustment This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 6 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 51: Enable/Disable Dose Calculation

    If reinstalling factory default parameters, all previously entered measured data (HVL val- ues and Dose exchange factors) will be lost. However, the measured data should already be noted in the document Radiographic Handbook. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 10 Medical Solutions Rev.
  • Page 52: Label

    The label is found in register 3 of the Technical Manual binder. affix the new label here mAs / mGy Program Auto Lim. Mo Rh WARNING Fig. 2 MAMMOMAT Novation SPB7-250.815.02 Page 2 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 53: Checking Opdose

    Repeat the procedure described in steps 2-5 for the other programs used. Record the measured values in the Startup Protocol, page 11 - 1. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 10 Medical Solutions Rev.
  • Page 54: Checking The Aec Dose Settings

    Mount the Collimator Mounted Plexi phantom and the compression plate simulator. Register or create a patient and submit with Exam. Fig. 3 Calibration - Patient register MAMMOMAT Novation SPB7-250.815.02 Page 4 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 55 The AEC sensors are drawn on the compression plate. Start an exposure. Measure the following values and add them to the table at step 4: MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 10 Medical Solutions Rev. 02 09.04...
  • Page 56 H and D Settings" and repeat the H & D dose check for Mo/Mo again. Record the measured values in the "Startup Protocol" on Page 11 - 1 . MAMMOMAT Novation SPB7-250.815.02 Page 6 of 10 Siemens AG System Rev. 02 09.04 CS SD 24...
  • Page 57: Check The H & D Dose Settings For Mo/Rh And W/Rh

    H and D Settings" and repeat the H & D dose checks for Mo/Rh and W/Rh again. Record the measured values in the "Startup Protocol" on Page 11 - 1 . MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 7 of 10 Medical Solutions Rev.
  • Page 58: Adjust The H And D Settings

    The BRICK software switches to a higher user level, thus a Username and Password have to be entered. Username: root Password: ****** MAMMOMAT Novation SPB7-250.815.02 Page 8 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 59 100,80,60 in the High Dose Scal- ing Value field and click Change High Dose Scaling to save it. Proceed from where you have been referenced from. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 9 of 10 Medical Solutions Rev.
  • Page 60 8 - 10 Checking dose settings This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 10 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 61: Checking Grid Lines

    The compression plate simulator is mounted on the MAMMOMAT. Checking visibility of grid lines Register or create a patient Grid lines test and submit with Exam. Fig. 1 Calibration - Patient register MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 20 Medical Solutions Rev. 02 09.04...
  • Page 62 Exposure Exp Mode mAs Tube Angle (yes/no) AEC "H" AEC "H" + 60 AEC "H" - 60 Tab. 1 Checking visibility of grid lines MAMMOMAT Novation SPB7-250.815.02 Page 2 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 63: Measuring Grid Lines

    Note the result in the Tab. 2. Choose the next image that had visible grid lines in the list and start over with step 3 until all images are evaluated. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 20 Medical Solutions Rev.
  • Page 64: Eliminating Grid Lines

    To minimize the risk that the high voltage is active when the grid is standing still, the time that the high voltage is disabled should be as long as possible. Move the grid to the left end position. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 65 9 - 5 Loosen the three screws. Now it's possible to adjust the grid switch (S1) alongside the grid (see Fig. 4). Fig. 4 Adjust grid switch S1 MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 20 Medical Solutions Rev. 02 09.04...
  • Page 66 A wheel is mounted on the left side of the grid motor (M1), on the wheel there is a slits which activates the grid switch (S1). Move the grid to the left end position. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 67 Open the carbon detector cover. Make sure the grid in mounted in parallel to the detector and lined up to the front of the table. The front switch has to be engaged. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 7 of 20 Medical Solutions Rev.
  • Page 68: Checking Image Quality

    Log into local service Options -> Service -> Local Service. NOTE The Service Level 1 or higher is necessary to do the Gain Calibra- tion. MAMMOMAT Novation SPB7-250.815.02 Page 8 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 69 9 - 9 Select Acquisition System. Fig. 8 Service Home Page As a result you will get the following screen. Fig. 9 Acquisition System Menu MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 9 of 20 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 70 The "temperature" field in this picture shows a value of "3045", due to a bug in the software. The actual temperature value is "30.45". MAMMOMAT Novation SPB7-250.815.02 Page 10 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 71 Press the Start button, to start the calibration. Fig. 11 Calibration - Start Create or register the patient Calibration and submit with Exam. Fig. 12 Calibration - Patient register MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 11 of 20 Medical Solutions Rev. 02 09.04...
  • Page 72 Check the image quality of the image. No objects or artifacts shall be visible in the image. If you Accept the image, the image will be used for Gain Calibration. MAMMOMAT Novation SPB7-250.815.02 Page 12 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 73 Procedure: Calibration Procedure and proceed with step 12 until you have a total of 8 images. Click Done to exit the Calibration. Close the examination. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 13 of 20 Medical Solutions Rev. 02 09.04...
  • Page 74: Creation Of New Pixel Map

    The Service Level 5 or higher is necessary to do the Bad Pixel Mapping. Select Acquisition System. Fig. 14 Service Home Page MAMMOMAT Novation SPB7-250.815.02 Page 14 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 75 As a result you will get the following screen. Fig. 15 Acquisition System Menu Click the Bad pixel mapping link. Fig. 16 Bad Pixel Mapping - Enter MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 15 of 20 Medical Solutions Rev. 02 09.04...
  • Page 76 Fig. 17 Bad Pixel Mapping - start If you want to end the Bad Pixel Mapping operation at this point, NOTE click the "Cancel" button. MAMMOMAT Novation SPB7-250.815.02 Page 16 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 77 Image Quality 9 - 17 Register the pre defined patient Calibration and submit with Exam. Fig. 18 Bad Pixel Mapping - Patient register MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 17 of 20 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 78 Image acquisition is in progress ..Saving processed Image ..Loading Image ..Click the Accept or Reject button. MAMMOMAT Novation SPB7-250.815.02 Page 18 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 79 NOTE collimation or misalignment or if there are any artifacts from debris or obstructions. Click Done to exit the Bad Pixel Mapping. Close the examination. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 19 of 20 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 80: Quality Control Manual Tests

    Make a note in the Startup Protocol page 11 - 1, that the Quality Control Manual tests have been performed MAMMOMAT Novation SPB7-250.815.02 Page 20 of 20 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 81: Settings Date And Time

    Go into the service software and select Utilities -> Escape to OS and enter the following command in the command field. start telnet brick A username and password have to be entered. Username: root Password: ****** MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 4 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 82 MM - minutes yy - year The following example sets the system to 3:10 p.m. on November 21, 2004: brick# date 1121031004 MAMMOMAT Novation SPB7-250.815.02 Page 2 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 83: Service Pc And Measuring Instruments

    (Mains input terminal) and all metallic parts of the equipment which the patient/operator may come in contact with, incl. all object tables. The protective ground resistance must not exceed 0.1 Ω. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 4 Medical Solutions Rev.
  • Page 84: Warning Label On Control Panel

    Installation or Start up problems occurred. Place a copy of the SP Q Quality Certificate in the technical documentation binder. • The Startup Protocol and the Installation Protocol remain with the system. MAMMOMAT Novation SPB7-250.815.02 Page 4 of 4 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 85: Site Data

    Luminance meter (measures the light intensity from the X-ray field) Dose meter Ion chamber Densitometer A non-invasive digital kV meter Tab. 1 Auxiliary materials MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 10 Medical Solutions Rev. 02 09.04 CS SD 24 System...
  • Page 86: Installation And Startup Steps Performed

    All Covers are mounted......... . . ❏ ❏ The temperature logger was sent to DirectRay Cooperation (DRC) ..MAMMOMAT Novation SPB7-250.815.02 Page 2 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 87 ........MAMMOMAT Novation Siemens AG SPB7-250.815.02...
  • Page 88: Startup Settings

    _____V x ( _____s + _____s + _____s) x 40 mA = ______ mAs 10 mAs Mo/Mo Tab. 2 Tube high voltage, current and mAs values MAMMOMAT Novation SPB7-250.815.02 Page 4 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 89 _____V x ( _____s + _____s + _____s) x 40 mA = ______ mAs Display ______ mAs W/Rh Tab. 2 Tube high voltage, current and mAs values MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 10 Medical Solutions Rev. 02 09.04 CS SD 24...
  • Page 90: Collimator Beam Calibration

    Results from the chapter "Checking the AEC dose settings" on Page 8 - 4. Anode / Dose value Pixel value Filter Mo/Mo 28 kV Mo/Rh W/Rh Tab. 5 AEC Dose Settings MAMMOMAT Novation SPB7-250.815.02 Page 6 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 91: Opdose Settings

    Angle (yes/no) ence [%] AEC "H" AEC "H" + 60 AEC "H" - 60 Tab. 7 Measuring Grid Lines MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 7 of 10 Medical Solutions Rev. 02 09.04 CS SD 24 System...
  • Page 92: Activity Report

    Number of employees who were employed in the period Total (Mh) Man hours, number of accumulated hours multiplied by number of employees MAMMOMAT Novation SPB7-250.815.02 Page 8 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 93: General Remarks

    __________________________________________________________________________ __________________________________________________________________________ A copy of this remains at the system and is filed in the Assembly/Installation folder. Place, date Installation team employee Siemens project manager/ person responsible for the project Name in block letters Name in block letters MAMMOMAT Novation Siemens AG SPB7-250.815.02...
  • Page 94 11 - 10 Startup Protocol This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 10 of 10 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 95: Troubleshooting Image Quality

    Isolation is the first step and should be done by using internally-generated image patterns in the image chain to determine where exactly the artifact is being produced (“what is/ what is not” the artifact analysis). MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 8 Medical Solutions Rev.
  • Page 96 Perform the bad pixel correction. ( Pixels and Lines ing. refer to "Creation of new pixel map" on Page 9 - 14) MAMMOMAT Novation SPB7-250.815.02 Page 2 of 8 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 97 Distorted image with Detector defective. Detector might have to be replaced 256 blocks contact the Service Center. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 3 of 8 Medical Solutions Rev. 02 09.04 CS SD 24 System...
  • Page 98: Working With The Service Pc

    Programmable Logic Device on the AEC board (D701). Flash Memory device on the AEC board (D701). ® ® Operating System, e.g. Microsoft Windows MAMMOMAT Novation SPB7-250.815.02 Page 4 of 8 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 99: Connecting The Service Pc

    If the IrDA-port was disabled in the computer BIOS, it should not be present in the list of available hardware. If the IrDA is present anyway and its icon is not marked with a red cross: Double-click on the device row. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 5 of 8 Medical Solutions Rev. 02 09.04...
  • Page 100: Starting Up And Using The Service Pc

    Choose Control Panel. Choose System. Choose the tab Hardware. Press the button Device Manager. Expand the row Ports. Double-click on the COM1 row. MAMMOMAT Novation SPB7-250.815.02 Page 6 of 8 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 101 12 - 7 Choose the tab Port Settings. Press the button Advanced. Uncheck the check box Use FIFO Buffers. Exit the dialog by pressing the OK button. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 7 of 8 Medical Solutions Rev. 02 09.04...
  • Page 102 12 - 8 Appendix This page intentionally left blank. MAMMOMAT Novation SPB7-250.815.02 Page 8 of 8 Siemens AG System Rev. 02 09.04 CS SD 24 Medical Solutions...
  • Page 103 Added time adjustment for the BRICK. Corrected the startup protocol regarding the opdima changes. Added a table for the auxiliary materials used with the installa- tion. MAMMOMAT Novation Siemens AG SPB7-250.815.02 Page 1 of 2 Medical Solutions Rev. 02 09.04...