Startup Settings; Tube High Voltage, Current And Mas Values - Siemens MAMMOMAT Novation DR Installation Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

11 - 4

Startup Settings

Tube high voltage, current and mAs values

Results from the chapter "Tube high voltage, current and mAs values" on Page 6 - 1.
kV
mAs
Anode/
filter
_____V x 5 kV =
30 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
20 mAs
Mo/Mo
_____V x 5 kV =
30 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
20 mAs
W/Rh
_____V x 5 kV =
35 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
100 mAs
Mo/Mo
_____V x 5 kV =
35 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
100 mAs
W/Rh
_____V x 5 kV =
25 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
100 mAs,
Mo/Mo
_____V x 5 kV =
25 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
100 mAs
W/Rh
_____V x 5 kV =
30 kV
_____V x ( _____s + _____s + _____s) x 40 mA =
10 mAs
Mo/Mo
Tab. 2
DR
MAMMOMAT Novation
System
Measured values
Tube high voltage, current and mAs values
SPB7-250.815.02
Rev. 02
09.04
Startup Protocol
Result
kV must be &5%
mAs must be &10%
______ kV
______ mAs
______ kV
______ mAs
______ kV
______ mAs
______ kV
______ mAs
______ kV
______ mAs
______ kV
______ mAs
______ kV
______ mAs
Page 4 of 10