Mechanical Checks
Mechanical Checks
Rotary motion
The rotary motion with the detector wing has to be checked.
1.
3
2
Fig. 6 Swivel arm system
2.
3.
Siemens AG
Medical Solutions
System OFF.
The rotary motion and the vertical adjustment of the swivel arm system will be
blocked.
1
System ON, detector table installed and projection angle set to minimum with
presetting knob (1/Fig. 6).
Check the rotary motion of the swivel arm system by pressing buttons (2/Fig. 6)
one by one.
The rotation of the swivel arm system will stop when the button is released or
when the end position is reached (+135 degrees clockwise and -180 degrees
counterclockwise). The projection angle is shown on the display at the lower part
of the stand.
4
2
3
SPB7-250.815.02
Rev. 02
09.04
MAMMOMAT Novation
Page 5 of 12
CS SD 24
5 - 5
5
5
DR
System