Siemens MAMMOMAT Novation DR Installation Manual page 59

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Checking dose settings
4.
Fig. 5 BRICK - Modify settings
5.
6.
Siemens AG
Medical Solutions
Click Modify settings.
To adjust the
- H value,
go to Modify low dose scaling (The default values are 80, 70, 45) and enter
the new values, separated by a comma, e.g. 80,70,45 in the Low Dose Scal-
ing Value field and click Change Low Dose Scaling to save it.
- D value,
go to Modify high dose scaling (The default values are 100, 80, 60) and enter
the new values, separated by a comma, e.g. 100,80,60 in the High Dose Scal-
ing Value field and click Change High Dose Scaling to save it.
Proceed from where you have been referenced from.
SPB7-250.815.02
Rev. 02
09.04
MAMMOMAT Novation
Page 9 of 10
CS SD 24
8 - 9