Vertical Adjustment - Siemens MAMMOMAT 1000 Nova Installation And Start-Up Instructions Manual

Hide thumbs Also See for MAMMOMAT 1000 Nova:
Table of Contents

Advertisement

15 - 2
Fig. 2 Display at the lower part of the stand
4. Set the projection angle to 60 degrees with presetting knob (2/ Fig. 1).
5. Press one of the buttons (1/  F ig. 1).
NOTE

Vertical adjustment

CAUTION
1. Press the buttons for upward and downward movement (3/  F ig. 1) of the swivel-
NOTE
M1000/3000 Nova
Installation and Start-Up
 o
 F
 O  C
The preset angle is displayed for approx 3 s.
The rotary motion of the swivel-arm system shall stop automatically when the
preset projection angle (both + and -) is reached and at 0 degrees.
All eight rotary motion buttons (four on either side) must be
checked. After the button has been released, the swivel-arm sys-
tem shall remain in the position set.
Remove the protective strips before performing vertical adjust-
ment of the swivel arm.
If not removed, the protective strips could be damaged.
Make sure the protective strips are removed before adjustments.,
see page 3 - 5.
arm system.
The movement shall stop automatically when either end position is reached or
when the button is released.
All eight vertical adjustment buttons (four on either side) must be
checked. The switching off in upper and lower end position is
effected by switches S881 and S884 respectively.
Register 3
Checking the swivel-arm system
 m  m
   6
   1
   2
   0
   0
   M
   A
   M
SPB7-230.033.11
Rev. 06
05.06
Page 2 of 4
CS PS SP
15
Siemens AG
Medical Solutions