MAMMOMAT 3000
Service
Repair Instructions
Replacing magnet for stereotactic device
Kit part No.;64 32 202 X041E
Register 5
Print No.: SPB7-230.091.07.01.02
© Siem ens-Elema AB 1 997
The reproduction, transmission or use of this
document or its contents is not permitted
without express written authority. Offenders will
be liable for damages. All rights, including
rights created by patent grant or registration of
a utility model
_or_ design,_are_
English
Doc. Gen. Date: 07.97
64 59 825
reserved.

Advertisement

loading

Summary of Contents for Siemens MAMMOMAT 3000

  • Page 1 MAMMOMAT 3000 Service Repair Instructions Replacing magnet for stereotactic device Kit part No.;64 32 202 X041E © Siem ens-Elema AB 1 997 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 Remove the bottom plate by unscrewing the eight remaining screws: - Two screws at either side of the biopsy unit. - Four screws at the bottom of the biopsy unit. Place the biopsy unit as shown in the picture. MAMMOMAT 3000 Register 5 RXB7-230.091.07 Page 2 of 4...
  • Page 3 Place the new magnet on the gliding plate and attach the cassette slider arm to the magnet. Reassemble the biopsy unit. Don´t forget to connect the grounding wire. Siemens-Elema AB Register 5 RXB7-230.091.07 Page 3 of 4 MAMMOMAT 3000 Solna, SWEDEN Page rev. 01/07.97 SPS-UM Service...
  • Page 4 This page intentionally left blank. MAMMOMAT 3000 Register 5 RXB7-230.091.07 Page 4 of 4 Siemen-Elema AB Service Page rev. 01/07.97 SPS-UM Solna, SWEDEN...