10

Preventive inspection sheet for illumination devices

· We recommend performing "Preventive Inspections" periodically (every time you replace lamps and at least once
every 6 months).
· The table below identifies the check items to be observed. Put (X) if not applicable or ( ) if applicable.
· If there are any check marks ( ), immediately stop using the product and request inspection to Olympus or replace
with new illumination device(s).
· If you detect an abnormality other than those listed below or Olympus products other than illumination devices, also
stop using the product and request inspection to Olympus.
· The repair, replacement and inspection that passed the warranty period are paid-services.
1.
More than 8 years have passed since original purchase of the illumination
device, or the total using time exceeds 20,000 hours of use.
2.
Lamp does not light sometimes even though the main switch is turned ON.
(Only when using the halogen bulb)
3.
Dirt or substances are attached around the main switch.
4.
The light flickers if you move a lamp cable or illumination devices.
5.
Lamp cable is unusually hot to the touch.
6.
Burning or smoke odor.
7.
The light still flickers even though the lamp is replaced.
(Only when using the halogen bulb)
8.
Signs of deformation, backlash, or looseness, etc. when you assemble/
disassemble the illumination device.
(I.e. it is hard to open/ close the lid during lamp-replacement.)
9.
Lamp connection terminals have become discolored or tarnished, or the
colors of the right and left terminals are different.
(Only when using the halogen bulb)
10. Illumination device/ housing has become deformed, cracked or tarnished in
any way.
11. Lamp cables or wiring parts have become deformed, cracked or tarnished
in any way.
12. Frequent repairs to similar devices put into use at the same time as the unit
being checked.
Check items
If you have any questions, please contact Olympus.
/
# If the spaces are not enough for check , copy this sheet.