Olympus BXFM Instructions Manual page 73

System microscope
Table of Contents

Advertisement

a
2
This picture shows the procedure to attach
BX3M-RLAS-S. The same procedure applies when
attaching other reflected light illuminators.
6
Attaching the reflected light illuminator
Attaching U-KMAS
1
b
2
a
Attaching the reflected light illuminator other than U-KMA
1
b
2
3
4
NOTE
Loosen the illuminator clamping screw a using the Allen screwdriver.
Insert the circular dovetail b at the lower area of the reflected light
illuminator for BF (U-KMAS) into the mounting portion of the illuminator
holder for BXFM-S (BXFM-ILHS) and tighten the clamping screw a
firmly.
Use the dedicated Allen wrench (
light illuminator to attach the reflected light illuminator.
Without aligning the mounting screws a (4 positions) of the reflected
light illuminator with the mounting screw holes b of the illuminator
holder for BXFM (BX3M-ILH) at first, place the reflected light illuminator
on the mounting surface at the slightly right position.
Then, push the reflected light illuminator to the end of the left side.
This is the correct attaching position.
While pushing the reflected light illuminator to the end of the left side,
tighten the mounting screws a (4 positions) using the dedicated
Allen wrench to secure the illuminator.
Attach the caps (4 positions) of the mounting screws provided with
the reflected light illuminator to the a position.
The cable comes out from the back of the coded reflected
light illuminator (BX3M-RLAS-S, BX3M-URAS-S). Be careful
not to put the cable between the illuminator and the
microscope frame.
Use the tweezers, etc when removing the caps of the screws
TIP
attached to the reflected light illuminator.
BXFM
) provided with the reflected
67