Olympus BXFM Instructions Manual page 82

System microscope
Table of Contents

Advertisement

7
76
6
Hold the new mercury burner wrapped with gauze, etc. and attach
the + (positive) pole of the mercury burner d to the fixed mount on
d
the upper side, then attach the - (negative) pole to the mount on the
lower side.
NOTE
7
Tighten the burner clamping screws (2 pcs.) of the socket part.
8
Align the external surfaces of the lamp housing with those on the
8
socket part, push the lamp housing straight downward and place the
socket part to the original position.
9
Tighten the clamping screw using the Allen screwdriver.
9
10
Set the life counter of the power supply for mercury burner (U-RFL-T) to
"0.0". For details, refer to the instruction manual provided with the unit.
USH-103OL: 300 hours
This value assumes the light cycles composed of 2 hours of
lighting and 30 minutes of extinction. Turning ON and OFF with
the cycle shorter than this will seriously shorten the service life of
the lamp.
Be careful not to touch the burner directly with your hand
and avoid leaving fingerprints or stains on the lamp.
Otherwise, the burner may be exploded due to the glass
distortion caused by the stains. If fingerprints or stains are
attached to the burner, clean it by wiping gently with a
piece of cleaning paper slightly moistened with absolute
alcohol.
Lamp service life