7. Repairguide for VENTImotion/VENTIlogic
7.1 General
Perform repairs on the VENTImotion or VENTIlogic exclusively at an ESD protected work-station!
•
Follow the safety rules in the operating instruc-
tions for the therapy device.
•
Any handling of the device assumes accurate
knowledge and observance of the operatingin-
structions and the service and repair guide.
•
Only perform repairs which are described in
this service and repair guide. This is the only
way to guarantee that the therapy device con-
tinues to function properly.
•
Ensure that your hands and your work-station
are clean during repair work.
7.2 Open device
Tools and equipment required
•
ESD protected work-station
•
3 mm Allen key
Caution!
Open the device only with the power supply plug disconnected.
1. Place the device on a non-slip surface with the
top facing downwards.
2. Remove the filter cassette from the top part of
the housing.
3. Undo and remove the 6 screws
•
•
•
.
34
Perform a function check after
(see "4. Test the device" on page 13).
If you replace components or individual parts,
use only original Weinmann parts.
When ordering the bottom part of the
housing
, quote the type, year of manufac-
35
ture and number of the device as well.
Note
The item numbers listed in the text which fol-
lows are identical to the item numbers in the
spare parts list on page 48 and in the over-
view on page 4.