Maintenance; Changing The Fuse, Dulcometer ® Compact Controller - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller ph/redox measured variable
Table of Contents

Advertisement

Maintenance

9
Maintenance
Users' qualification: trained user, see
n
Ä Chapter 1.2 'Users' qualifications'
on page 7
The DULCOMETER
is maintenance free.
9.1
Changing the fuse, DULC‐
OMETER
troller
WARNING!
Danger from electrical voltage
Possible consequence: Fatal or very
serious injuries.
The DULCOMETER
Controller does not have a mains
switch
When working inside the control
unit, disconnect the control unit
from the mains power via an
external switch or by removing
the external fuse
NOTICE!
Use only 5 x 20 mm micro-fuses
Possible consequence: Damage to
the product or its surroundings
5x20 T 0.315 A
Part number 732404
80
®
Compact Controller
Compact Con‐
®
Compact
®
Fuse change
The mains fuse is located in a sealed fuse
holder in the inside of the device.
1.
Disconnect the controller from the
mains power
2.
Open the controller and fold the
controller housing top section to the
left
3.
Remove the PCB cover
4.
Remove the micro-fuse using a
suitable tool
5.
Fit the micro-fuse using a suitable
tool
6.
Refit the PCB cover
7.
Replace controller housing top sec‐
tion and close the controller