Devices' Technical Parameters; General Features - Innomed Medical Hear Screen 112C-1 Service Manual

Table of Contents

Advertisement

Innomed Medical

3 Devices' technical parameters

Classification according to 93/42/EEC directive:

3.1 General features

Reference temperature range
Operating temperature range
Humidity
Shipping/storage temperature range
Operating environment
Operating and storage pressure
Environment abstinence
Mains and power supply
Mains voltage
Mains frequency
Mains power consumption
Accumulator type
Accumulator nominal voltage
Accumulator charging time (informative)
Accumulator charger type, protection
Operating time from accumulator
Safety
Safety class
Patient protection (with PC connection as
well)
Operating modes
Monitoring, display
Manual printing
Automatic printing:
Record storage (STORE)
Analysis and diagnosis creation
(DGN)
Arrhythmia analysis (ARTM)
Heart Screen 112C-1
II.a
This product has CE mark according to the 93/42/EEC Council
Directive on medical devices, dated on 14. June 1993.
Intelligent quick charger, thermal protection at 54 C, thermal fuse at
60 minutes of ECG curve printing in intermittent operating mode, or
3, 6 or 12-channel monitoring + rhythm channel
Manual printing from any lead groups on 3, 6 or 12 channels
Adjustable length automatic printing from the 12 standard ECG leads in 2x6
or 4x3 channel groups.
Automatic recording and storage from 12 leads. Rhythm curve may be stored
in pre-programmed mode.
Automatic 12-channel recording with analysis and diagnosis. Raw curve,
rhythm curve may be stored in pre-programmed mode. Data, marked average
curved appearance on the display and printer.
Arrhythmia event storage, trends, statistics.
Result appearance on the printer and display.
20...95% humidity (without condensation)
IP20, the device protected from middle-sized solid objects
(>12mm) but it is not protected from water intrusion
2A@110VAC 1 A@230VAC
only Innomed produced accumulator is useable.
In case of totally dead accumulator max. 2 hours
8 hours of monitoring (only basic functions are operated)
4/47
Service manual
0... +40 C
0... +45 C
-20... +55C
dry, closed space
550 – 1060 hPa
110-230VAC
50 - 60 Hz
2100 mAh,
9,6 V
90 C
I.
Defibrillator protected CF
D-7443-1-C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents