Device Level Measurements; Heartscreen 80G-L Basic Functions - Innomed Medical Heart Screen 60G Service Manual

Simultaneous 12-channel ecg devices with 58 / 80mm wide 1- / 3-channel built-in printer and lcd display
Hide thumbs Also See for Heart Screen 60G:
Table of Contents

Advertisement

Innomed Medical

8 Device level measurements

8.1

HeartScreen 80G-L Basic functions

1.
Connect a device top part without the printer head and the generator.
2.
Before downloading the program connect the power supply set to +9V in place of the battery. Remove the
resistor.
3.
Connect the other power supply – set to 3.6V – to the J17 battery drain point. Connect the negative points of
the two power supplies.
Test procedure
Backlight test
Amplifier part test
Memory test
Keyboard test
Speaker test
RTC (clock) test
Contrast test
Burn impulse test
Serial data transfer test
Software Low Batt. check
Checking power off
Note:
To enter the test menu, press and hold the Menu/Mon and Speed buttons when turning the device on.
2014.01.09
Heart Screen 60G-80G-L
Comment
Backlight is operational when the device is on.
In the test menu.
In the test menu.
In the test menu.
In the test menu.
Set the exact date and time, which has to be the same after turning the card
off and on again.
The LCD contrast is acceptable, not foggy, the LCD is operational.
Start from the test menu.
Max. 5ms burn time measured at J11.8.
(Active signal at high level)
ECG signal monitoring with Innobase for Windows program and generator,
checking main functions.
Gradually decrease battery voltage and with it the ONPOWER voltage
level (the processor checks this every second), at 8.4V +/-0.1V in the top
left corner a symbol appears:
(ONPOWER J11.9)
Continue to decrease the battery voltage (the processor checks this every
second), the device turns off at 7.2V +/-0,1V (ONPOWER J11.9).
32/84
Service manual
D-5450-1-B_HS60G80GL_SM_ENG.doc