3-2 Service
Return Door Solenoid
3. Remove the
using a 5/16" nut driver. Do not completely remove the nuts.
Figure 3-2. Loosening the nuts holding the metal plate
4. Maneuver the metal mount around the sheet metal tabs and nuts until the lower rim can exit
the housing first. The solenoid is connected by a length of yellow wire long enough to allow
pulling the PCB/solenoid metal plate out of the ERB.
Figure 3-3. Removing the metal mount
5. Remove the two
Figure 3-4. Removing the solenoid screws
6. Unplug the solenoid at the 2-pin Molex connector.
7. Remove the defective
8. Plug in the 2-pin Molex connector.
External Return Bin Installation and Service Guide/67-2038 Rev D
PCB/solenoid metal plate
6-32 PH Torx screws
solenoid
and replace it.
by loosening the two nuts on either side of the plate,
using a Torx screwdriver #10 and set them aside.