6. Functional Check

6.1 Intervals

6.2 Method

Carry out a functional check on the unit at least every
6 months.
If you find a defect during the functional check,
SOMNOsmart 2 should not be reused until the defect
has been remedied.
1. Assemble SOMNOsmart 2 with the hose system
, exhalation system
12
supply cable
2. Press the on/off switch
softstart is switched on, briefly press the softstart
button
to switch softstart off. The radial fan
25
delivers air through the hose system to the mask,
and the current pressure in hPa is shown in the LC
display
.
21
3. Close the opening of the exhalation system, e.g.
with your thumb or hand.
Compare the pressure shown in the LC display with
the prescribed minimum pressure (see patient pass).
Deviations should not be more than max. ±1 hPa.
4. Switch the unit off with the button
If the values or functions specified here are not fulfilled,
please send the unit for repair to the dealer or
manufacturer.
, mask
13
ready for operation.