1.2 Function

SOMNOsmart 2 works on the principle of an
electrically powered flow generator, and delivers an
electronically controlled constant pressure level.
A radial fan draws in ambient air through a filter
and delivers it to the outlet port. From here the air
flows through the hose system and the mask to the
patient.
An oscillation generator creates a slight pneumatic
sinusoidal oscillation in the air flow produced by
the device.
The airway and oscillation pressure is registered
via the pressure measurement tube in the mask
The current oscillation pressure reading is com-
pared with an average value and the preceding
values. This evaluation of the oscillation pressure
makes it possible to identify any significant
change in airway resistance and hence to in-
crease or reduce the therapy pressure via the flow
generator.
The pressure present in the mask is shown in the
LC display. The microprocessor-controlled flow
generator regulates the pressure at the value
determined.
The device also records a respiratory flow signal
that can be output to a PSG system or read off with
the aid of WEINMANNsupport. This makes it
possible to identify apnoea, hypopnoea and
respiratory flow limitations.
The expiration system before the mask prevents ac-
cumulation of CO
system.
-enriched expired air in the tube
2
Description
5
EN