8. Before calling the service technician (Warning)
module id message id
10045
10048
19
Specified End Date is earlier than Start Date.
Length of acquired worklist item is too long.
The following tag number item(s) are cut off.
20
[XXXX] (DICOM TAG Number+DICOM TAG
Name)
Acquired worklist item format is invalid.
The following tag number item(s) are not acquired.
21
[XXXX] (DICOM TAG Number+DICOM TAG
Name)
1
Invalid Values for Birth Date
Patient Info. set as any of essential data items are
5
not entered.
The length of either of 5 Patient Name fields is too
6
long.
Shorten the length to 60 characters or less.
12
Invalid character was entered to the field.
Line feed code cannot be entered to Study
14
Description field.
Character '\' or control character cannot be used for
22
Accession No.
Character '\' or control character cannot be used for
23
Patient ID.
Character '\,' line feed code, or control character
24
cannot be used for Study Description.
Character '\', '=' or control code cannot be used for
29
Patient Name.
Multi-byte character cannot be used for Patient
30
Name.
Out-of-Range Birth Date (A date earlier than 1869
39
or later than tomorrow is entered.)
40
Entered date is invalid for Birth Date.
A character/characters for Patient ID is/are not
42
supported by Specific Character Set.
A character/characters for Accession No. is/are not
43
supported by Specific Character Set.
A character/characters for Patient Name is/are not
44
supported by Specific Character Set.
A character/characters for Study Description is/are
45
not supported by Specific Character Set.
A character/characters for the field is/are not
46
supported by Specific Character Set.
A character/characters for the field is/are not
47
supported by Specific Character Set.
A character/characters for the field is/are not
48
supported by Specific Character Set.
Entered characters are not acceptable because they
49
do not match Specific Character Set.
Length of the acquired item is too long.
50
The following item(s) is/are cut off.
Message
-122-
Detail
Input a date later than the Start Date as the End Date.
The transfer data of the worklist server does not
meet the DICOM standards. Check the worklist
server settings.
Input the correct birth date.
Input the patient information that was set in the
essential data (see page 83).
Shorten the length to 60 characters or less.
Check the details of the displayed item, and enter it
again.
Delete the line feed code.
Delete the backslash (\) and any control characters.
Delete the backslash (\), line feed code and any
control characters.
Delete the backslash (\) and any control characters.
Delete the multi byte characters.
Input the correct birth date.
Input using the specific character set.
Delete the single-byte Japanese kana characters.
Check the information input in the magnetic card or
barcode.