Canon CR-1 MARK II Operation Manual page 121

Retinal imaging control software nm
Hide thumbs Also See for CR-1 MARK II:
Table of Contents

Advertisement

module id message id
1001
Failed to open a port.
1003
Failed to set a port.
Power for the retinal camera is turned off or the
1007
cable is improperly connected.
Error occurred during communications with the
1501
retinal camera.
Failed to resend the command frame.
Error occurred during communications with the
10005
1502
retinal camera.
Failed to resend the reply frame.
Error occurred during communications with the
1503
retinal camera.
Time-out for No Response
Error occurred during communications with the
1504
retinal camera.
Time-out for Incomplete Transmission
Cameras are not attached to the main mount and
1603
sub-mount of the retinal camera.
Connected retinal camera model is different from
1701
the model setting in this software.
10007
4
JPEG File Saving Error
Failed to log in.
1
Confirm User ID/Password.
10010
2
It cannot be set without administrator's privileges.
2
This application is already running.
101
Set the screen resolution to 1024 × 768 or more.
102
Set the screen color to 24 bits or more.
10011
Operation can be executed by a Windows user
110
with Administrator privileges.
Essential data has not been set up.
114
Set up essential data.
Free space in the drive is not sufficient to capture
images.
Make up space in the drive by deleting studies in
27
Study Logs List.
[Drive:XX freesize: XXGByte(XXbyte)]
[Drive: XX]
Image cannot be captured due to insufficient space
10012
to store captured image.
29
Make up space in the drive by deleting studies in
Study Logs List.
[Drive:XX freesize: XXGByte(XXbyte)]
Number of captured images exceeds the maximum
30
for the study.
No more image can be captured for the study.
7
MPPS Communication Error
10019
Study cannot be executed since a study having the
11
same Study Instance UID is running.
8. Before calling the service technician (Error)
Message
-113-
Detail
Refer to 7.3.1 Retinal Camera Settings (see
page 77), and check the port number setting.
Disconnect the retinal camera communication
cable, then connect the cable again.
Refer to 7.3.1 Retinal Camera Settings (see
page 77), and check the port number setting.
Check that the computer and retinal camera are
connected correctly.
Turn the retinal camera's power off and back on.
Check that the digital camera is connected correctly.
Check the model of the retinal camera.
Check how much free memory is left on the drive
where the data is to be saved.
Input the User ID and Password registered in the
user account (see page 100), and log in.
Log in using the User ID for which the Privilege has
been set to the Administrator (see page 100).
Use the software application which is running.
Set the Windows screen resolution to 1024 × 768 or
more.
Set the Windows color quality to 24 bit or more.
Log into Windows using the name of the user
belonging to the Administrators group.
Input the required items.
Delete studies in the study log screen (see
page 38) to increase the amount of available space
in the image storage drive.
End the study, and start a new study.
Check that the MPPS server is operating properly.
Check the MPPS tab (see page 91) settings.
Check the study information which was transmitted
from the server.
Continued on the following page >>