Chapter 3. Device Setting
Pace Detect: Available options are Off and On. Electrocardiograph system detects and rejects
pacemaker-generated signals from ECG signal so that they will be ignored in determining heart
rate. If you select On for patients with pacemaker, detected pacemaker signals will be marked
on the ECG waveform as a white vertical line.
For patients with pacemaker, the PACE DETECT function must
be switched "ON", otherwise, the pacing impulse may be counted as
normal QRS complex.
Next page: By pressing this button from User Setting menu, the window depicted below
will be displayed.
Periodic Recording: Press to set time interval in periodic recording. Available options are
5-60 min and Off. If you select Off, periodic recording will be stopped.
Periodic Interval Repetition: Press to set repetition of recording. Available options are 1-
20 and Infinite.
Print info: Select on to record header of the printed paper and select off to remove it.
Measurement Lead: Future capability.
Previous page: Press to come back to User Setting Menu page.
Figure 3-9 second page of user setting