 External surfaces
In-between patients and as required:
For cleaning: wipe gently using a moist cloth and warm soapy water or mild detergent and for
disinfection use the following recommended agents:
■ Alcohol 70%
■ Isopropyl alcohol
■ N-propanol
 Display screen
In-between patients and as required use clean and soft cloth with neutral detergent and with
Isopropyl alcohol may be used for cleaning and disinfection.
1) Take extra care when cleaning the screen because it is more sensitive
to rough cleaning methods than the housing.
2) Don't spray a liquid directly on the screen.
 Accessories
To clean, disinfect and sterilize reusable ECG cable, lead wires & electrodes (Clamp electrode,
Chest electrode, Suction chest electrode), and so forth, refer to the instructions delivered with
the accessories.
Also, trolley (if applicable) should be cleaned and disinfected after each patient or when
necessary, using a soft, clean cloth soaked in mild soapy water and, if necessary, Isopropyl
alcohol, and then wiped with a dry cloth.
1) To avoid damaging of the ECG cable, lead wires and electrodes do not immerse it in
any liquid.
2) Disposable accessories shall not be sterilized or reused.
3) To prevent environmental pollution, the disposal of single-use accessories shall be
done in accordance with the policies of the hospital.
 Recorder
Accumulation of paper powder or foreign matter between the thermal head and platen roller
deteriorates the print quality. Clean the head elements and platen roller surface using alcohol
and a cotton swab. Wait until the alcohol dries then close the recorder door.
Do not clean the recorder immediately after recording because thermal head and
its surrounding area are hot during and after recording.