Measured Peak Excitation Transients For 1000 Foot Lengths Of Three Belden Lead Wires Used By Campbell Scientific - Campbell Measurement and Control Module CR10 Operator's Manual

Ds 70;
Hide thumbs Also See for Measurement and Control Module CR10:
Table of Contents

Advertisement

SECTION 13. CR10 MEASUREMENTS
TABLE 13.3-4. Measured Peak Excitation Transients for 1000 Foot Lengths of Three Belden
Vx(mV)
8641
2000
1000
NOTE: Excitation transients are eliminated
if excitation leads are contained in a shield
independent from the signal leads.
The size of the peak transient is linearly related
to the excitation voltage and increases as the
bridge resistor, R
, increases. Table 13.3-4
f
shows measured levels of V
lengths of three Belden wires used in Campbell
Scientific sensors. Values are given for R
equal to 1 kohm and 10 kohm. Table 13.3-4 is
meant only to provide estimates of the size of
excitation transients encountered; the exact
level will depend upon the specific sensor
configuration.
Equation 13.3-7 can be solved for the
maximum lead length, L, permitted to maintain
a specified error limit. Combining Equations
13.3-7 and 13.3-4 and solving for L gives:
L = -(R
C
+ (t/ln(V
/V
o
f
e
eo
where V
is the measurement error limit.
e
EXAMPLE LEAD LENGTH CALCULATION
FOR 107 TEMPERATURE SENSOR
Assume a limit of 0.05°C over a 0°C to +40°C
range is established for the transient settling
error. This limit is a reasonable choice since it
approximates the linearization error over that
range. The output signal from the thermistor
bridge varies nonlinearly with temperature
ranging from about 100 µV/=°C at 0°C to 50
µV/°C at 40°C. Taking the most conservative
figure yields an error limit of V
other values needed to calculate the maximum
lead length are summarized in Table 13.3-5
and listed below:
13-8
Lead Wires Used by Campbell Scientific
-----------------------V
R f =1 kohm
#
#
8771
50
100
25
65
for 1000 foot
eo
f
)))/R
C
[13.3-15]
o
w
= 2.5 µV. The
e
(mV) -----------------------
eo
#
#
8723
8641
60
100
40
60
1) V
~ 50 mV, peak transient at 2 V excitation
eo
2) V
~ 2.5 µV, allowable measurement error
e
3) t = 450 µs, CR10 input settling time
4) R
= 1 kohm, 107 probe source resistance
o
5) C
= 3.3 nfd, CR10 input capacitance
f
6) C
~ 42 pfd/ft., lead wire capacitance
w
Solving Equation 13.3-15 gives a maximum
lead length of:
L ~ 1003 ft., error ~ 0.05°C
Setting the allowable error at 0.1°C or
approximately 5 µV, the maximum lead length
increases to:
L ~ 1085 ft., error ~ 0.1°C
13.3.4 SUMMARY OF SETTLING ERRORS FOR
CAMPBELL SCIENTIFIC RESISTIVE
SENSORS
Table 13.3-5 summarizes the data required to
estimate the effect of lead length on settling
errors for Campbell Scientific's resistive
sensors. Comparing the transient level, V
the input range, one suspects that transient
errors are the most likely limitation for the 107
sensor. The sensors in the WVU-7 are the
same as in the Model 107 (the lead wire is
different), but the signal leads for the WVU-7
wet- and dry-bulbs are not subject to excitation
transients because they are shielded
independently from the excitation.
R
=10 kohm
f
#
#
8771
8723
140
80
90
40
, to
eo

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds 70

Table of Contents

Save Article as PDF