Wiring Diagram For Prt In 4 Wire Half Bridge - Campbell Measurement and Control Module CR10 Operator's Manual

Ds 70;
Hide thumbs Also See for Measurement and Control Module CR10:
Table of Contents

Advertisement

FIGURE 7.9-1. Wiring Diagram for PRT in 4 Wire Half Bridge
The result of Instruction 9 when the first
differential measurement (V
the 2.5 V range is equivalent to R
Instruction 16 computes the temperature (°C)
for a DIN 43760 standard PRT from the ratio of
the PRT resistance at the temperature being
measured to its resistance at 0°C (R
/R
Thus, a multiplier of R
f
9 to obtain the desired intermediate, R
(=R
/R
x R
/R
). If R
and R
s
f
f
o
s
exactly 100 ohms, the multiplier would be 1.
However, neither resistance is likely to be
exact. The correct multiplier is found by
connecting the PRT to the CR10 and entering
Instruction 9 with a multiplier of 1. The PRT is
then placed in an ice bath (@ 0°C; R
the result of the bridge measurement is read
using the *6 Mode. The reading is R
is equal to R
/R
since R
o
f
value of the multiplier, R
this reading. The initial reading assumed for
this example was 0.9890. The correct multiplier
is: R
/R
= 1/0.9890 = 1.0111.
f
0
The fixed 100 ohm resistor must be thermally
stable. Its precision is not important because
the exact resistance is incorporated, along with
that of the PRT, into the calibrated multiplier.
SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
) is not made on
1
/R
.
s
f
/R
).
s
0
is used in Instruction
0
/R
s
0
were each
0
=R
), and
s
0
/R
, which
s
f
=R
. The correct
s
o
/R
, is the reciprocal of
f
0
The 10 ppm/°C temperature coefficient of the
fixed resistor will limit the error due to its
change in resistance with temperature to less
than 0.15°C over the specified temperature
range. Because the measurement is
ratiometric (R
/R
), the properties of the 10
s
f
kohm resistor do not affect the result.
PROGRAM
01:
P9
Full BR w/Compensation
01:
1
Rep
02:
23
25 mV 60 Hz rejection EX
Range
03:
23
25 mV 60 Hz rejection BR
Range
04:
1
IN Chan
05:
1
Excite all reps w/EXchan 1
06: 2100
mV Excitation
07:
1
Loc [:Rs/Ro
08:
1.0111
Mult (Rf/Ro)
09:
0
Offset
02:
P16
Temperature RTD
01:
1
Rep
02:
1
R/Ro Loc Rs/Ro
03:
2
Loc [:TEMP C ]
04:
1
Mult
05:
0
Offset
]
7-7