Settling Error, In Degrees, For 024A Wind Direction Sensor Vs. Lead Length; Resistive Half Bridge Connected To Single-Ended Cr10 Input - Campbell Measurement and Control Module CR10 Operator's Manual

Ds 70;
Hide thumbs Also See for Measurement and Control Module CR10:
Table of Contents

Advertisement

FIGURE 13.3-6. Resistive Half Bridge Connected to Single-Ended CR10 Input
R
, the source resistance, is not constant
o
because R
varies from 0 to 10 kohms over the
b
0 to 360 degree wind direction range. The
source resistance is given by:
R
= R
+(R
(R
-R
+R
)/(R
o
d
b
s
b
f
R
+(R
(20k-R
)/20k)
d
b
b
Note that at 360 degrees, R
of 6k (R
=10k) and at 0 degrees, R
b
(R
=0). It follows that settling errors are less at
b
lower direction values.
The value of R
for any direction D (degrees) is
b
given by:
R
(kohms) = (10k)(D)/360
b
Equation 13.3-6 can be rewritten to yield the
settling error of a rising signal directly in units of
degrees.
Error (degrees) = De -t/(R
Equation 13.3-12, -13 and -14 can be combined
to estimate the error directly in degrees at
various directions and lead lengths (Table 13.3-
3). Constants used in the calculations are
given below:
C
= 3.3nfd
f
C
= 41 pfd/ft., Belden #8771 wire
w
t = 450µs
+R
)) =
s
f
[13.3-12]
is at a maximum
o
is 1k
o
[13.3-13]
(C
+C
L))
o
f
w
[13.3-14]
SECTION 13. CR10 MEASUREMENTS
TABLE 13.3-3. Settling Error, in Degrees,
for 024A Wind Direction Sensor vs. Lead
Wind
Direction
360°
270°
180°
90°
The values in Table 13.3-3 show that significant
error occurs at large direction values for leads
in excess of 500 feet. Instruction 4, Excite,
Delay, and Measure, should be used to
eliminate errors in these types of situations.
Using a 10 ms delay, settling errors are
eliminated up to lengths that exceed the drive
capability of the excitation channel (~ 2000 ft.).
13.3.3 TRANSIENTS INDUCED BY SWITCHED
EXCITATION
Figure 13.3-6 shows a typical half bridge
resistive sensor, such as Campbell Scientific's
Model 107 Temperature Probe, connected to
the CR10. The lead wire is a single-shielded
pair, used for conducting the excitation (V
signal (V
) voltages. When V
s
transient is capacitively induced in V
signal voltage. If the peak transient level, V
is less than the true signal, V
has no effect on the measurement. If V
greater than V
, it must settle to the correct
so
signal voltage to avoid errors.
Length
Error
L=1000 ft.
L=500 ft.
66°
15°
45°
21°
) and
x
is switched on, a
x
, the
s
, the transient
so
is
eo
,
eo
13-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds 70

Table of Contents

Save PDF