FIGURE 7.17-1. CR10/Paroscientific "T" Series Transducer Wiring Diagram
Subroutine 1, which loads the coefficients into
input locations, is called only on the first
execution following program compilation.
The temperature frequency is read on single-
ended Channel 12 and pressure is measured
on single-ended Channel 1.
Temperature, T
, D and C are computed in
0
Subroutine 2 using a generalized fourth order
polynomial equation. The form of the equation
is
F(x) = U
+ U(M
+ U(M
0
1
This form provides faster execution and greater
accuracy of the floating point math. "M
represents coefficients 0 through 4, and U is
defined above. With five coefficients, T
only parameter requiring a fourth order
polynomial. To maintain the fourth order
format, zero value "dummy" coefficients are
plugged in where required for the temperature,
D, and C coefficients.
Instructions to output the readings to Final
Storage are not included in this example.
SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
+ U(M
+ UM
))).
2
3
4
"
0..4
is the
0
*
1
Table 1 Programs
01:
60
Sec. Execution Interval
01:
P89
If X<=>F
01:
42
X Loc CMPILE CK
02:
1
=
03:
0
F
04:
1
Call Subroutine 1
02:
P27
Period Average (SE)
01:
1
Rep
02:
4
Input gain=1
03:
12
IN Chan
04: 9000
No. of Cycles
05:
6
Time Out (units=.01sec)
06:
10
Loc [:UT us
07:
1
Mult
08:
0
Offset
03:
P27
Period Average (SE)
01:
1
Rep
02:
3
Input gain=10
03:
1
IN Chan
04: 5000
No. of Cycles
05:
16
Time Out (units=.01sec)
06:
11
Loc [:TAU us ]
07:
1
Mult
08:
0
Offset
]
7-21