Hook Up To Avw1 - Campbell Measurement and Control Module CR10 Operator's Manual

Ds 70;
Hide thumbs Also See for Measurement and Control Module CR10:
Table of Contents

Advertisement

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
Program: AVW1 & CR10 USED TO
MEASURE 1 GEOKON VIBRATING WIRE
SENSOR.
*
1
Table 1 Programs
01:
60
Sec. Execution Interval
01:
P4
Excite,Delay,Volt(SE)
01:
1
Rep
02:
15
2500 mV fast Range
03:
1
IN Chan
04:
1
Excite all reps w/EXchan 1
05:
1
Delay (units .01sec)
06: 2500
mV Excitation
07:
1
Loc [:TEMP
08:
.001
Mult
09:
0
Offset
7-18
FIGURE 7.16-3. Hook up to AVW1
02:
03:
]
P55
Polynomial
01:
1
Rep
02:
1
X Loc TEMP
03:
1
F(X) Loc [:TEMP
04:
-104.78
C0
05:
378.11
C1
06:
-611.59
C2
07:
544.27
C3
08:
-240.91
C4
09:
43.089
C5
P28
Vibrating Wire (SE)
01:
1
Rep
02:
2
IN Chan
03:
1
Excite all reps w/EXchan 1
04:
24
Starting Freq. (units=100 Hz)
05:
32
End Freq. (units=100 Hz)
06:
500
No. of Cycles
07:
0
Rep delay (units=.01sec)
08:
2
Loc [:PRESSURE ]
09:
-34.836
Mult
]

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds 70

Table of Contents