Ohm Prt In 3 Wire Half Bridge; Wire Half Bridge Used To Measure 100 Ohm Prt - Campbell Measurement and Control Module CR10 Operator's Manual

Ds 70;
Hide thumbs Also See for Measurement and Control Module CR10:
Table of Contents

Advertisement

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
FIGURE 7.10-1. 3 Wire Half Bridge Used to Measure 100 ohm PRT
7.10 100 OHM PRT IN 3 WIRE HALF
BRIDGE
The temperature measurement requirements in
this example are the same as in Section 7.9. In
this case, a three wire half bridge, Instruction 7,
is used to measure the resistance of the PRT.
The diagram of the PRT circuit is shown in Fig.
7.10-1.
As in the example in Section 7.9, the excitation
voltage is calculated to be the maximum
possible, yet allow the +25 mV measurement
range. The 10 kohm resistor has a tolerance of
±1%; thus, the lowest resistance to expect from
it is 9.9 kohms. We calculate the maximum
excitation voltage (V
across the PRT less than 25 mV:
0.025V > V
x
V
The excitation voltage used is 2.1 V.
The multiplier used in Instruction 7 is
determined in the same manner as in Section
7.9. In this example, the multiplier (R
assumed to be 100.93.
The 3 wire half bridge compensates for lead
wire resistance by assuming that the resistance
of wire A is the same as the resistance of wire
B. The maximum difference expected in wire
resistance is 2%, but is more likely to be on the
order of 1%. The resistance of R
with Instruction 7, is actually R
difference in resistance of wires A and B. The
average resistance of 22 AWG wire is 16.5
7-8
) to keep the voltage drop
x
115.54/(9900+115.54);
< 2.17 V
x
/R
) is
f
0
calculated
s
plus the
s
ohms per 1000 feet, which would give each 500
foot lead wire a nominal resistance of 8.3 ohms.
Two percent of 8.3 ohms is 0.17 ohms.
Assuming that the greater resistance is in wire
B, the resistance measured for the PRT (R
100 ohms) in the ice bath would be 100.17
ohms, and the resistance at 40°C would be
115.71. The measured ratio R
the actual ratio is 115.54/100 = 1.1554. The
temperature computed by Instruction 16 from
the measured ratio would be about 0.1°C lower
than the actual temperature of the PRT. This
source of error does not exist in the example in
Section 7.9, where a 4 wire half bridge is used
to measure PRT resistance.
The advantages of the 3 wire half bridge are
that it only requires 3 lead wires going to the
sensor and takes 2 single-ended input
channels, whereas the 4 wire half bridge
requires 4 wires and 2 differential channels.
PROGRAM
01:
P7
3 Wire Half Bridge
01:
1
Rep
02:
23
25 mV 60 Hz rejection Range
03:
1
IN Chan
04:
1
Excite all reps w/EXchan 1
05: 2100
mV Excitation
06:
1
Loc [:Rs/Ro
07:
100.93
Mult
08:
0
Offset
02:
P16
Temperature RTD
01:
1
Rep
02:
1
R/Ro Loc Rs/Ro
03:
2
Loc [:TEMP C ]
04:
1
Mult
05:
0
Offset
=
0
/R
is 1.1551;
s
0
]

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds 70

Table of Contents

Save PDF