0-trAC
ZERO TRACKING [1306]
This function allows you to perform temporary zero calibration compensating for the temperature drift
of the weight.
Switching off the transmitter automatically restores the previous zero calibration.
The maximum weight that can be reset by this parameter is 2% of the capacity of the system.
To disable this function, set the value 0.
Value
Variation
0
Control excluded
1
0.5 div/sec
2
1 div/sec
3
2 div/sec
4
3 div/sec
Default: 0
0-bAnd
ZERO BAND [1307]
This parameter defines the number of divisions that can be reset by pressing the zero button on the
front or the associated Input.
Values: from 0 to 200.
Default: 100