Troubleshooting
7.2
There is nothing to see on the LCD screen: The
battery is almost empty.
Still images or videos cannot be recorded: There
is no SD card in the device.
Off: The battery is not charged enough.
None of the functions work or hang.
If the functions cannot be restored:
Reset
7.3
Locking may be fully or partially engaged.
Articulation cables may be constricted by bends in
the flexible probe.
The probe tip may be blocked because of their
position within the device.
Do not apply excessive force on control levers.
NOTICE
> Damage or destruction of the probe tip
Top-Line Video Endoscope
Video Display are not Working Properly
Articulation is Difficult
Recharge the battery.
Insert the SD card into the video endoscope.
Charge the battery.
Switch off the video endoscope and restart the
device.
Remove the cover from the battery compart-
ment. Remove the battery to reveal the reset
button.
Insert a toothpick or another non-conducting
tool into the reset hole and push the reset but-
ton to perform a system reset.
Release locking.
Keep the flexible probe as straight as possible
during inspection.
Reposition the probe tip by rotating the flexible
probe or moving the probe tip forwards and
backwards.