Dimensional Drawings ........................... 23 Electrical Installation ......................28 Process Interface ............................29 Process Interface thermoMETER TIM 8 / thermoIMAGER TIM 41 ..............29 Process Interface thermoIMAGER TIM 40 ..................... 32 Pin Assignment thermoMETER TIM 8 / thermoIMAGER TIM 41 ..............34 Pin Assignment thermoIMAGER TIM 40 .......................
Page 4
Cleaning ................................. 37 Software TIM Connect ......................38 Minimum System Requirements ........................38 Installation and Initial Operation ........................38 Autonomous Operation thermoMETER TIM 8 / thermoIMAGER TIM 41 ............41 9.3.1 Special Settings for thermoIMAGER TIM 41 ................43 9.3.2 Hot-/Coldspot Function in Autonomous Operation ..............
Page 5
Emissivity Table Metals ......................84 Emissivity Table Non Metals ....................87 Industrial Process Interface for thermoMETER TIM 8 (Optional) ........89 Industrial Process Interface for thermoIMAGER TIM 40 (Optional) ........93 Examples for a Fail-Safe Monitoring with a PLC ..............96 USB Cable Extension for thermoIMAGER TIM 40 ..............
> Irreparable defect of the infrared detector Avoid static electricity, arc welders, and induction heaters. Keep away from very strong EMF (electromagnetic fields). > Damage to or destruction of the camera thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 7...
European standards (EN). The measuring system is designed for use in industrial environ- ments. The EU Declaration of Conformity and the technical documentation are available to the responsible authori- ties according to the EU Directives. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 8...
Technical Data Technical Data Functional Principle The thermoMETER TIM 8 and the thermoIMAGER TIM 40 calculates the surface temperature based on the emitted infrared energy of objects, see 10. The two-dimensional detector (FPA - focal plain array) allows a measurement of an area and will be shown as thermographic image using standardized palettes. The radio- metric processing of the picture data enables the user to do a comfortable detailed analysis with the software TIM Connect.
IEC 68-2-6: (sinus shaped), see 2.4 IEC 60068-2-64 (broadband noise), see 2.4 Shock IEC 68-2-27: 25 g and 50 g, see 2.4 1) Used standards for vibration and shock thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 11...
Shock, half sinus 50 g – testing Ea 50 g (acc. IEC 60068-2-27) Acceleration 490 m/s (50 g) Pulse duration 11 ms Number of directions (3 axes with 2 directions each) Duration 18 shocks (3 shocks each direction) thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 12...
Page 13
150 - 500 Hz 1.9230 (m/s (0.020 g /Hz) 500 - 2000 Hz -6 dB/ octave 2000 Hz 0.1245 (m/s (0.00126 g /Hz) Number of axes Duration (3 x 1 h) thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 13...
Flag Status, alarm, frame sync, fail-safe, external communication cess Interface Digital interface USB 2.0/ Ethernet/ RS485 USB 2.0/ optional USB to GigE USB 2.0/ Ethernet/ RS485 (PoE) conversion thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 14...
80 mK (NETD) Warm-up time 10 min Emissivity 0.100 ... 1.100 Software TIMConnect 1) Accuracy statement effective from 150 °C 2) At ambient temperature 23±5 °C; whichever is greater. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 15...
Delivery Delivery Unpacking, Included in Delivery 1 thermoMETER TIM 8, thermoIMAGER TIM 40 or thermoIMAGER TIM 41 1 USB cable (1 m 1 Ethernet PoE cable: 1 m (only for thermoIMAGER TIM 41) 1 Mounting nut and mounting bracket (adjustable in one axis, tripod thread)
Different parameters are important if using infrared cameras. They display the connection between the dis- tance of the measured object and the size of the pixel, see Fig. 3, see Fig. 4, see Fig. 5. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 17...
Page 18
MFOV Recommended, smallest measured object size of 3 x 3 Pixel (thermoIMAGER TIM 40 / TIM 41) and 2 x 2 (thermoMETER TIM 8), not shown in the drawing above, see Fig. 2. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41...
Page 19
TIM Connect has an algorithm which corrects this distortion. As an alternative to the tables below, the optics calculator can also be used. You will find the latest software at: https://www.micro-epsilon.com/temperature-sensors/thermoIMAGER/optics/ thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 19...
VFOV: Vertical enlargement of the total measuring field at object level IFOV: Size of the single pixel at object level 1) The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 20...
VFOV: Vertical enlargement of the total measuring field at object level IFOV: Size of the single pixel at object level 1) The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 21...
VFOV: Vertical enlargement of the total measuring field at object level IFOV: Size of the single pixel at object level 1) The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 22...
Mechanical Installation The thermoMETER TIM 8 and the thermoIMAGER TIM 40 / TIM 41 are equipped with a metric M30x1 thread and can be installed either directly via the sensor thread or with help of the supplied mounting nut (standard) and adjustable mounting bracket (standard) to a mounting device available, see Fig.
Page 24
Mechanical Installation Fig. 8 Dimensional drawing thermoMETER TIM 8, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 24...
Page 25
Mechanical Installation Fig. 9 Dimensional drawing thermoMETER TIM 40, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 25...
Page 26
Mechanical Installation Fig. 10 Dimensional drawing thermoIMAGER TIM 41, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 26...
Page 27
Mechanical Installation Fig. 11 Industrial/stackable PIF (Process Interface) - electronic box, control box shutter, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 27...
Electrical Installation Electrical Installation At the back side of the thermoMETER TIM 8 / thermoIMAGER TIM 41 and the thermoIMAGER TIM 40 cam- eras there are the two connector plugs, see Fig. 12, see Fig. 13. Fig. 12 Back side of the thermoMETER TIM 8 / Fig.
Process Interface thermoMETER TIM 8 / thermoIMAGER TIM 41 The thermoMETER TIM 8 / thermoIMAGER TIM 41 is equipped with an integrated process interface (cable with terminal block included in scope of supply), which can be programmed via the software as an direct ana-...
Page 30
Electrical Installation Fig. 14 Configuration terminal block thermoMETER TIM 8 / thermoIMAGER TIM 41 Designation Description Color Shield Black Ground Brown 5 - 30 V Power supply White Analog/Digital Input or RS485 (A) Green Analog output or RS485 (B) Yellow...
Page 31
Electrical Installation The thermoMETER TIM 8 / thermoIMAGER TIM 41 provides the following direct inputs and outputs: Designation Description max range / status Analog input 0 - 10 V Digital input 24 V (Low-active = 0 ... 0.6 V) Analog output...
Main measure area, measure area, internal temperature, flag status, alarm, frame sync, fail-safe, external communication Digital input (DI) Flag control, triggered recording, triggered snapshots, triggered line-scanner, triggered event grabber, reset peak-/valley-hold thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 32...
Page 33
1) Depending on supply voltage; for 0 - 10 V on the AO the PIF has to be powered with min. 12 V. 2) The AI is designed for max. 24 V, the voltage level above 10 V is not interpreted. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 33...
Pin Assignment thermoMETER TIM 8 / thermoIMAGER TIM 41 Ethernet RS485 or AO RS485 or AI GND-ISO Fig. 16 Rear side thermoMETER TIM 8 / thermoIMAGER TIM 41 connector thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 34...
In this case, avoid a voltage > 3 V at INT pin. > Destruction of the camera 1) We recommend using only a switching contact between INT and DGND as external hardware (button, relay). thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 35...
Adjust the focus of the image by using the distance function in the software (menu View > Windows > Distance or over the icon Fig. 20 Window Distance thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 36...
(e.g. Purosol or B+W Lens Cleaner). Never use cleaning compounds which contain solvents (neither for the lens nor for the housing). > Destruction of the sensor and/or the controller thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 37...
5. Choose the desired language in the menu Tools > Language. 6. Adjust the focus of the image by using the distance function in the software (Menu View > Win- dows > Distance or over the icon thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 38...
Page 39
Software TIM Connect Fig. 21 Window Distance Fig. 22 Software window thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 39...
Page 40
Status bar: Serial number, optic, temperature range, cursor position, device framerate/ display frame- rate, emissivity, ambient temperature, flag status Further details to the software are in the enclosed USB stick. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 40...
Autonomous Operation thermoMETER TIM 8 / thermoIMAGER TIM 41 A special feature of the thermoMETER TIM 8 / thermoIMAGER TIM 41 is the autonomous operation. There is no need for a permanent connection to the TIM Connect software. Only a few settings must be set in advance in the software.
Page 42
By pressing the OK button an @ sign can be found in the configuration menu Devices (PIF), which indicates the autonomous operation. Fig. 24 Configurations menu Device (PIF) Fig. 25 Setup thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 42...
The resulting value can be displayed for example on a multimeter, see Fig. When using this variant, it can last for about half a minute until the device is ready for use in autonomous mode. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 43...
Page 44
Software TIM Connect Fig. 27 Electrical installation for autonomous operation of thermoMETER TIM 8 / thermoIMAGER TIM 41 thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 44...
Page 45
Software TIM Connect - Autonomous operation also works via the industrial/stackable PIF of the thermoMETER TIM 8 / ther- moIMAGER 41. The device is powered by the power supply of the PIF. - Up to 9 measure areas (for TIM 8 with firmware 3013 or higher) and up to 3 measure areas (TIM 41) can be output autonomously.
Note: Predefined Layout in the software available under Tools and Layouts: TIM 8 Hot spot autonomous or TIM 41 Hot spot autonomous. Fig. 29 Measure area over entire field of view thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 46...
Ethernet thermoMETER TIM 8 / thermoIMAGER TIM 41 The thermoMETER TIM 8 / thermoIMAGER TIM 41 has a direct Ethernet interface. The advantage is cable lengths of up to 100 m. For example, a switch can be used to extend the distances.
IP address camera: 192.168.0.101 IP address PC: 192.168.0.100 Port number: 50101 To do this, go to Control Panel and open the Network and Sharing Center. Go on LAN connection. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 48...
Page 49
(192.168.0.100). This must be identical to the address set in the TIM Connect software. Then close the windows with OK. The network settings on the PC are complete. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 49...
Page 50
The approval of programs can also be activated afterwards in the Windows Firewall settings of the PC (under Windows Fire- wall and Allow a Program or feature through Windows Firewall). thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 50...
Page 51
Listen on port number. The port range used should be between 49152 ... 65535. When using other ports, it could be that they are already reserved or assigned. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 51...
- Different layout options for an individual setup (arrangement of windows, toolbar) - Range of individual measurement parameter fitting for each application - Adaption of thermal image (mirror, rotate) - Individual start options (full screen, hidden, etc.) thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 52...
Page 53
- Definition of visual or acoustic alarms and analog data output - Analog and digital signal input (process parameter) - External communication of software via Comports, DLL and LabVIEW - Adjustment of thermal image via reference values thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 53...
Page 54
- Data with color information for standard programmes such as Photoshop or Windows Media Player (*.avi, *.jpg, *.tiff) - Data transfer in real time to other software programmes via LabVIEW, DLL or Comport interfaces thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 54...
The temperature rose even more in the area behind the red end of the spectrum. Finally he found the maximum temperature far behind the red area. Nowadays this area is called infrared wavelength area. Fig. 31 William Herschel (1738 – 1822) thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 55...
Page 56
Distance to Spot size. The spectral filter selects the wavelength range, which is relevant for the temperature measurement. The detector in cooperation with the processing electronics transforms the emitted infrared radiation into electrical signals. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 56...
Page 57
- very fast response and exposure times - measurement without inter-reaction, no influence on the - measuring object - non-destructive measurement - long lasting measurement, no mechanical wear Fig. 34 TIM with tablet PC thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 57...
Development of elec- cabinets ing plastic parts tronic components Monitoring of cables Development of me- Process control at calen- Process control in solar chanical parts dering cell production thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 58...
Fig. 35 Composition of IR radiation IR radiation Reflection + + = 1 Emission Transmission thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 59...
Page 60
Emissivity Fig. 36 Spectral emissivity of several materials Concrete Enamel Plaster Chamotte thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 60...
Cover a part of the surface of the measuring object with a black, flat paint with an emissivity of 0.98. Adjust the emissivity of your infrared thermometer to 0.98 and take the temperature of the colored surface. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 61...
Page 62
Afterwards, determine the temperature of a directly adjacent area and modify the emissivity until the measured value corresponds to the temperature of the colored surface. On all three methods the object temperature must be different from ambient temperature. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 62...
- Spectral range of the measurement - Transmissivity (e.g. with thin films) Fig. 40 Adjustment of the emissivity in the software TIM Connect under the menu item Tools / Configuration / Device thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 63...
MICRO-EPSILON or to your distributor / retailer. MICRO-EPSILON undertakes no liability whatsoever for damage, loss or costs caused by or related in any way to the product, in particular consequential damage, e.g., due to...
Raleigh, NC 27617 /USA Tel. +1 919 / 787-9707 Fax +1 919 / 787-9706 me-usa@micro-epsilon.com www.micro-epsilon.com For customers in Canada or South America applies: Please contact your local distributor. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 65...
- A list of national laws and contacts in the EU member states can be found at https://ec.europa.eu/environment/topics/waste-and-recycling/waste-electrical-and-electronic-equipment-weee_en. Here you can inform yourself about the respective national collection and return points. - Old devices can also be returned for disposal to MICRO-EPSILON at the address given in the imprint at https://www.micro-epsilon. de/impressum/.
The air purge can be mounted in four different positions. The direction of the airflow must always be clear. Fig. 41 Directions of airflow at different mounting positions thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 67...
Page 68
Appendix | Accessories Fig. 42 Laminar air purge with SI protective window TM-APL-TIM8/40, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 68...
Page 69
Appendix | Accessories Fig. 43 Laminar air purge with SI protective window TM-APL and mounting bracket TM-APLAB, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 69...
Page 70
Appendix | Accessories Fig. 44 Mounting bracket TM-APLAB, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 70...
Then using water cooling, a corresponding mounting kit (TM-WAKx-TIMxx) is required. Water flow rate: approx. 2 l/ min (cooling water temperature should not exceed 30 °C) Fig. 45 Water cooling TM-W-TIM8/4 and mounting kit TM-WAK1-TIMxx, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 71...
Page 72
Appendix | Accessories Fig. 46 Shutter for water cooling TM-SW-TIM8/40, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 72...
S4 on one control box must be set to mA and the others to mV, see Fig. Ambient temperature: 0 ... +60 °C (T camera: 0 ... +50 °C); with water cooled housing up to +250 °C thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 73...
Page 74
Appendix | Accessories Fig. 47 Shutter TM-S-TIM8/40 with mounting bracket TM-APLAB-TIM8/40, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 74...
Page 75
Appendix | Accessories Fig. 48 Shutter TM-S-TIM8/40, dimensions in mm thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 75...
Page 76
IN 1: Trigger input for normal operation (S1) 24 V, input is active LOW (open IN 2: Currently no usage input = HIGH)) IN 3: Trigger input for fast-closing mode (S2) thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 76...
Water cooling and shutter TIM 8 Water cooling and shutter TIM 40 Air purge, water cooling and shutter TIM 8 Air purge, water cooling and shutter TIM 40 thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 77...
Page 78
Fig. 50 Air purge TM-APL, water cooling TM-W-TIM8/40, shutter TM-SW-TIM8/40 and appropriate mounting kit TIM-WAKx-TIMxx, dimensions in mm Components - Air purge - Water cooling - Shutter - Mounting kit thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 78...
Page 79
Appendix | Accessories Fig. 51 Air purge TM-APL, water cooling TM-W-TIM8/40 and appropriate mounting kit TIM-WAKx-TIMxx, dimen- sions in mm Components - Air purge - Water cooling - Mounting kit thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 79...
Page 80
Appendix | Accessories Fig. 52 Water cooling TM-W-TIM8/40L, shutter TM-SW-TIM8/40 and appropriate mounting kit TIM-WAKx-TIMxx, dimensions in mm Components - Water cooling - Shutter - Mounting kit thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 80...
Page 81
Appendix | Accessories Fig. 53 Air purge TM-APL-TIM8/40 with mounting bracket TM-FB-TIM8/40 and shutter TM-S-TIM8/40, dimen- sions in mm Components - Air purge - Mounting bracket - Shutter thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 81...
TM-OPH24-TIM8/40. In addition, the industrial or stackable PIF TM-CJAPIF-TIM (thermoIMAGER TIM 40) or TM-PIFCx-TIM8 (thermoMETER TIM 8 / thermoIMAGER TIM 41) can be installed as an accessory without housing and a USB Server TM-USBSGB-TIM. Fig. 54 Outdoor protective housing for TIM camera...
The devices have following presettings at time of delivery: Temperature range -20 ... 100 °C Emissivity 1.000 Process interface (PIF) inactive Interprocess Communication (IPC) inactive Measurement function Rectangle measure area thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 83...
For use in industrial environment the industrial process interface with 500 V AC isolation voltage between thermoMETER TIM 8 and process is available (connection box with IP65, 5 m, 10 m or 20 m standard or high temperature cable for camera connection, terminal block for process integration), see A 12.
Page 90
Appendix | Industrial Process Interface for thermoMETER TIM 8 (Optional) The industrial process interface provides the following inputs and outputs: Designation Description max. range / status INPUT 1/2/3 Analog- or digital input 1, 2, 3 0 - 10 V AO 1/2/3...
Page 91
Appendix | Industrial Process Interface for thermoMETER TIM 8 (Optional) TIM 8 Fig. 56 Connection of 3 industrial PIFs via RS485 thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 91...
Page 92
Appendix | Industrial Process Interface for thermoMETER TIM 8 (Optional) Each stackable industrial PIF must have its own RS485 address. The address must be set directly on the board and in the software. For the PIF which is the furthest away the 120R switch (TERM.- Termina- tion) has to be set.
Pin assignment Color Designation Grey Interrupt Green SCL (I Yellow SCL (I White 3.3 V Brown Shield Fig. 57 Connections of the industrial Process Interfaces for thermoIMAGER TIM 40 thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 93...
Page 94
2) The AI is designed for max. 24 V, the voltage level above 10 V is not interpreted. 3) Active if AO1, 2 or 3 is/ are programmed as alarm output. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 94...
Page 95
Interruption data cable camera - Interruption power supply PIF Shut-down of TIM Connect software Crash of TIM Connect software Fail-Safe-Output 0 V at analog output (AO) Open contact (fail-safe-relay)/ green LED off thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 95...
[4] Malfunction of camera [2] Cable break of fail-safe cable [5] Breakdown of TIM power supply/ Interruption of USB cable [3] Interruption of cable TIM-PIF [6] Malfunction of TIM Connect software thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 96...
Page 97
[2] Cable break of fail-safe cable [6] Breakdown of TIM power supply/ Interruption of USB cable [3] Short circuit of fail-safe cable [7] Malfunction of TIM Connect software [4] Interruption of cable TIM-PIF thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 97...
The maximum USB cable length is 20 m. For greater distances between thermoIMAGER TIM 40 and com- puter or for stand-alone solutions the optional USB Server Gigabit is provided: Fig. 60 USB-Server Gigabit thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 98...
Command List The command list is provided on the USB stick and the TIM Connect software (Help > SDK). Every command must expire with CR / LF (0x0D, 0x0A). thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 99...
The main difference to the former Version 1 (ImagerIPC.dll) is the support of more than one TIM cameras via multiple instances of TIM Connect. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 100...
To limit this effort and to enable anybody to translate the resources of the TIM Connect ap- plication Micro-Epsilon has developed the small tool Resource Translator. This tool helps to translate any visible text within the thermoIMAGER TIM Connect application. thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 101...
The digital input can be activated with a switch to the TIM GND or with a Low level CMOS/TTL signal: - Low level 0 … 0.6 V - High level 2 … 24 V Fig. 62 Button example thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 102...
Appendix | Process Interface for thermoIMAGER TIM 40 A 12.3 Analog Input Fig. 63 Analog input Useful voltage range: 0 ... 10 V thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 103...
The voltage level for AO1 - AO3 can be set in the software: - no alarm: 0 V - Alarm: 2 – 10 V REL1-3 (D01-D03): U = 30 VDC = 400 mA Fig. 64 Relay output thermoMETER TIM 8 / thermoIMAGER TIM 40 / TIM 41 Page 104...
Page 106
MICRO-EPSILON MESSTECHNIK GmbH & Co. KG Koenigbacher Str. 15 · 94496 Ortenburg / Germany Tel. +49 (0) 8542 / 168-0 · Fax +49 (0) 8542 / 168-90 X9751390-B042092HDR info@micro-epsilon.com · www.micro-epsilon.com Your local contact: www.micro-epsilon.com/contact/worldwide/ MICRO-EPSILON MESSTECHNIK...