Maintenance ■ Troubleshooting
Yellow
Red
Red/Yellow
Red
Red
- 132 -
• Tray not present
• Z MOTION ERROR
800: LIMIT
EXCEEDED ON AXIS 2
• Error Tape Out
• Main Air OFF
• Vacuum error
PSV5000 does not detect the tray and when
you click Run, you get error message Tray Not
Present; or the Run window shows a tray
where there is none.
• Re-install the tray.
• At the Run window, the Tray Present Sensor
should be ON.
• If the tray is still not detected, remove the
tray and place a finger over tray present
sensor.
–If sensor is working, tray outline appears.
Re-install tray.
–If sensor is not working, the tray outline
does not appear—contact Data I/O
Customer Support or a local Data I/O
approved service representative
Conditions: A Red ERROR box appears in
bottom right corner of the Monitor.
Contact Data I/O Customer Support or a local
Data I/O approved service representative.
• If a device is jammed in the Tape Output
System, clear it. Acknowledge the error
message by pressing the ESC key. Then click
OK on the PSV5000 Monitor.
• If above doesn't work, call the Service
Technician.
• Check that the main input air valve is in the
ON position.
• Check that the main air hose is connected.
• Check that the air pressure is within the
required range (green LED readout instead of
red.)
• Contact Data I/O Customer Support or a
local Data I/O approved service
representative.
• See Adjusting the Vacuum Generator Sensor
on page 105.
• Contact Data I/O Customer Support or a
local Data I/O approved service
representative.