(10) Specify the resistor box's setting to 350 ohm.
(11) Press the [START] key.
(12) When you have finished the 350-ohm Setup operation, the
Press the [ENTER] key to begin writing the new Setup data to the EEPROM.
(13) After the setup parameters have been registered to the EEPROM, the display closes and returns to
normal mode.
(14) Turn off the GL450/400/350.
This completes the RTD temp. setup.
GL400-UM-251-9370
"EEPROM :
4-10
4. SETUP PROCEDURES
[ENTER]" message appears.