Kodak X-Omat 480 RA Processor Installation Instructions Manual page 37

Hide thumbs Also See for X-Omat 480 RA Processor:
Table of Contents

Advertisement

Installing the Replenishment Tubes Through Alternate Inlets
Prevent contaminating the processing solutions. Do the following procedure to connect the
DEVELOPER REPLENISHMENT TUBE to the DEVELOPER REPLENISHMENT TANK, then, do
the same procedure to connect the FIXER REPLENISHMENT TUBE.
[1]
Insert the REPLENISHMENT TUBE through one of the holes in the bottom of the PROCESSOR. The
tubing is not supplied.
Only thick-walled TUBING will withstand the high suction of the REPLENISHMENT PUMPS.
[2]
Install a GROMMET onto the REPLENISHMENT TUBE. If necessary, remove a section of the
GROMMET.
[3]
Compress the GROMMET and insert it into the hole in the bottom of the PROCESSOR.
Figure 44 Using the Alternate Inlets for the Replenishment Tubes
1C0932
NOTE
ALTERNATE
INLETS
H108_0235ACA
H108_0235AA
37

Advertisement

Table of Contents
loading

Table of Contents