Kodak X-Omat 480 RA Processor Installation Instructions Manual page 35

Hide thumbs Also See for X-Omat 480 RA Processor:
Table of Contents

Advertisement

[4]
Connect the DUCT to the PROCESSOR.
DUCT from
PROCESSOR
Figure 41 Connecting the Processor Exhaust to the Building Exhaust
Connecting the Exhaust Fan for Continuous Operation
The PROCESSOR is shipped with the exhaust fan wired for intermittent operation while the
PROCESSOR is energized. Intermittent has proven to be the more economical mode of
operation and causes the least amount of wear on the PROCESSOR. For the exhaust fan to
operate continuously, do the following:
[1]
Move the wall power switch to the ''OFF''
position.
[2]
Move CB1 to the ''O'' position.
[3]
Remove the DRYER PANEL and RECEIVING
END ACCESS PANEL.
[4]
Open the ELECTRICAL BOX COVER.
[5]
Disconnect wire 20-25 from TB4-4 and
connect it to TB4-3.
[6]
Close the ELECTRICAL BOX COVER.
[7]
Install the DRYER PANEL and RECEIVING
END ACCESS PANEL.
[8]
Move the wall power switch to the ''ON''
position.
[9]
Move CB1 to the ''I'' position.
1C0932
Adjust clearance to obtain
required static pressure in duct.
Do not make solid connection.
5.1 cm (2.0 in.)
maximum
NOTE
Figure 42 Wiring for Continuous Operation
of the Exhaust Fan
H108_0008BCB
H108_0008BA
H108_0244AA
35